A clinical evaluation of the Terumo Capiox SX18R hollow fiber oxygenator.
نویسندگان
چکیده
The Terumo Capiox SX18R is a commercially available, low prime, reverse phase, hollow fiber membrane oxygenator. The oxygenator consists of a 1.8 m2 microporous polypropylene hollow fiber bundle, a 2200 cm2 tubular stainless steel heat exchanger, and an open hard shell venous reservoir with integral cardiotomy filter. The Terumo Capiox SX18R oxygenator was evaluated to determine its clinical oxygenating performance. Blood samples were drawn from 25 patients yielding 114 data points. The following parameters were recorded: blood flow, cardiac index, gas flow, gas to blood flow ratio, and oxygen fraction. Samples were assayed for hematocrit, hemoglobin, arterial and venous blood gas values, and venous oxygen saturation. The data and assay results were used to calculate arterial, venous, and membrane gas oxygen content, oxygen transfer, shunt fraction, and oxygen diffusion capacity. The Terumo Capiox SX18R oxygenator performed adequately with sufficient oxygen transfer reserve and carbon dioxide clearance under a variety of clinical conditions for the tested population.
منابع مشابه
Clinical evaluation of the Terumo Capiox FX05 hollow fiber oxygenator with integrated arterial line filter.
Perfusion techniques and equipment in pediatric open heart surgery have continued to focus on decreasing prime volumes and lowering surface areas of the cardiopulmonary bypass circuit. While this has improved drastically over the last 20 years, greater demand is being placed on the perfusionist to reduce the deleterious effects of bypass without compromising safety or efficiency. Specifically, ...
متن کاملClinical gaseous microemboli assessment of an oxygenator with integral arterial filter in the pediatric population.
The use of an arterial line filter (ALF) within the pediatric cardiopulmonary bypass (CPB) circuit is not a new concept. It has always presented the perfusionist with a circuit component that while valuable, increased prime volume. The purpose of this study was to evaluate the change in prime volume and emboli between a conventional oxygenator with separate ALF and a new generation oxygenator w...
متن کاملDisconnection of Cobe SMARxT tubing from the venous outlet of the Terumo Capiox SX25RX oxygenator during cardiopulmonary bypass.
The use of surface modified, biocompatible tubing in cardiopulmonary bypass has been reported to decrease the inflammatory responses caused by blood contact with the non endothelial surface of poly vinyl chloride (PVC) tubing. The combination of advances in biocompatible tubing and increased affordability resulted in a change to our cardiopulmonary bypass circuit, with the Terumo Capiox SX25 ox...
متن کاملNon-destructive evaluation of membrane lung gas exchange performance.
This paper describes a method of evaluating the gas exchange effectiveness of hollow fiber oxygenators utilizing gas on both sides of the membrane. The goal of the study was to develop an evaluation technique which was accurate, reliable, and did not harm or contaminate a new, sterile oxygenator. Three pediatric oxygenators were tested and compared: the Medtronic Minimax Plus, the Terumo Capiox...
متن کاملRandomized trial of the Terumo Capiox FX05 oxygenator with integral arterial filter versus Terumo Capiox Baby RX05 and Terumo Capiox AF02 arterial filter in infants undergoing cardiopulmonary bypass.
UNLABELLED The purpose of this clinical trial was to evaluate the effect of the Terumo Capiox FX05 oxygenator with integrated arterial filter during cardiopulmonary bypass (CPB) compared with the Terumo Capiox RX05 Baby RX and arterial filter on inflammatory mediators and blood product utilization. Forty patients weighing less than 10 kg who underwent congenital heart surgery utilizing cardiopu...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The journal of extra-corporeal technology
دوره 27 3 شماره
صفحات -
تاریخ انتشار 1995