Avoidance of radiation injuries from medical interventional procedures.
نویسنده
چکیده
Interventional radiology (fluoroscopically-guided) techniques are being used by an increasing number of clinicians not adequately trained in radiation safety or radiobiology. Many of these interventionists are not aware of the potential for injury from these procedures or the simple methods for decreasing their incidence. Many patients are not being counselled on the radiation risks, nor followed up when radiation doses from difficult procedures may lead to injury. Some patients are suffering radiation-induced skin injuries and younger patients may face an increased risk of future cancer. Interventionists are having their practice limited or suffering injury, and are exposing their staff to high doses. In some interventional procedures, skin doses to patients approach those experienced in some cancer radiotherapy fractions. Radiation-induced skin injuries are occurring in patients due to the use of inappropriate equipment and, more often, poor operational technique. Injuries to physicians and staff performing interventional procedures have also been observed. Acute radiation doses (to patients) may cause erythema at 2 Gy, cataract at 2 Gy, permanent epilation at 7 Gy, and delayed skin necrosis at 12 Gy. Protracted (occupational) exposures to the eye may cause cataract at 4 Gy if the dose is received in less than 3 months, at 5.5 Gy if received over a period exceeding 3 months. Practical actions to control dose to the patient and to the staff are listed. The absorbed dose to the patient in the area of skin that receives the maximum dose is of priority concern. Each local clinical protocol should include, for each type of interventional procedure, a statement on the cumulative skin doses and skin sites associated with the various parts of the procedure. Interventionists should be trained to use information on skin dose and on practical techniques to control dose. Maximum cumulative absorbed doses that appear to approach or exceed 1 Gy (for procedures that may be repeated) or 3 Gy (for any procedure) should be recorded in the patient record, and there should be a patient follow-up procedure for such cases. Patients should be counselled if there is a significant risk of radiation-induced injury, and the patient's personal physician should be informed of the possibility of radiation effects. Training in radiological protection for patients and staff should be an integral part of the education for those using interventional techniques. All interventionists should audit and review the outcomes of their procedures for radiation injury. Risks and benefits, including radiation risks, should be taken into account when new interventional techniques are introduced.A concluding list of recommendations is given. Annexes list procedures, patient and staff doses, a sample local clinical protocol, dose quantities used, and a procurement checklist.
منابع مشابه
Comprehensive TLD dosimetry of physicians and patients during interventional angiography procedures to estimate the risk of radiation deterministic effect
Introduction: Interventional angiography procedures are increasingly used today. These techniques require a long time of standing near the patient and x-ray tube which lead to increase the radiation-dose of the staff and interventionists. This may also, in turn, lead to the occurrence of radiation deterministic effects. Since a large number of interventional radiology were perf...
متن کاملPATIENT SAFETY Patient Radiation DoseManagement in the Follow-Up of Potential Skin Injuries in Neuroradiology
BACKGROUND AND PURPOSE: Radiation exposure from neurointerventional procedures can be substantial, with risk of radiation injuries. We present the results of a follow-up program applied to potential skin injuries in interventional neuroradiology based on North American and European guidelines. MATERIALS ANDMETHODS: The following guidelines approved in 2009 by SIR and CIRSE have been used over t...
متن کاملEvaluation of Maximum Patient Skin Dose Arising from Interventional Cardiology Using Thermoluminescence Dosimeter in Mashhad, Iran
Introduction The increasing practice of interventional fluoroscopy in diagnosis and treatment of cardiovascular disease has risen attention to improve radiation protection of patients and cardiologists in these relatively high dose techniques. Therefore, nowadays there is an emphasis on the measurement of radiation dose received by patients and cardiologists arising from the relevant procedures...
متن کاملPatient dose values during interventional cardiology examinations in Yazd hospital, Iran
Background: The number of interventional cardiology (IC) procedures has increased rapidly. coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) are now widely performed as a matter of routine, and they are considered safe procedures for experienced cardiologists. However, it is also known that these procedures are associated with high radiation doses due to long f...
متن کاملRadiation-induced skin injuries from fluoroscopy.
Since 1992, the U.S. Food and Drug Administration (FDA) has received reports of radiation-induced injuries to the skin in patients who had undergone fluoroscopically guided interventional procedures. The reports were investigated to determine the procedure- or equipment-related factors that may have contributed to the injury. The injuries ranged in severity from erythema to moist desquamation t...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Annals of the ICRP
دوره 30 2 شماره
صفحات -
تاریخ انتشار 2000