Comparison of trigeminal neuralgia radiosurgery plans using two film detectors for the commissioning of small photon beams
نویسندگان
چکیده
Trigeminal neuralgia (TN) is a chronic, episodic facial pain syndrome that can be extremely intense, and it occurs within the regions of the face that are innervated by the three branches of the trigeminal nerve. Stereotactic radiosurgery (SRS) is the least invasive procedure to treat TN. SRS uses narrow photon beams that require high spatial resolution techniques for their measurement. The use of radiographic or radiochromic films for small-field dosimetry is advantageous because high spatial resolution and two-dimensional dose measurements can be performed. Because these films have different properties, it is expected that the calculated dose distributions for TN patients will behave differently, depending on the detector used for the commissioning of the small photon beams. This work is based on two sets of commissioned data: one commissioned with X-OMAT V2 film and one commissioned with EBT2 film. The calculated dose distributions for 23 TN patients were compared between the commissioning datasets. The variables observed were the differences in the half widths of the 35 and 40 Gy isodose lines (related to the entrance distance to the brainstem) and the volume of the brainstem that received a dose of 12 Gy or more (V12). The results of this comparison showed that there were statistically significant differences between the two calculated dose distributions. The magnitudes of these differences were up to 0.33 mm and 0.38 mm for the 35 and 40 Gy isodose lines. The corresponding difference for the V12 was up to 2.1cc. It is clear that these differences may impact the treatment of TN patients, and then it must be important to perform this type of analysis when observing complication rates. Clinical reports on irradiation techniques for trigeminal neuralgia should consider that different detectors used for commissioning treatment planning systems might result in small but significant differences in dose distributions.
منابع مشابه
Measurements of the relative output factors for CyberKnife collimators.
OBJECTIVE To determine accurately the relative output factors, defined as the ratio of the nominal dose rate for a given collimator to that of the 60-mm collimator. This is particularly important for radiosurgical treatment of functional disorders, such as trigeminal neuralgia, in which a single large radiation dose is delivered to the target with a small collimator, such as the 5-mm collimator...
متن کاملSmall photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملEvaluation of MRI-based MAGIC polymer gel dosimeter in small photon fields
Background: Accurate small radiation field dosimetry is essential in modern radiotherapy techniques such as stereotactic radiosurgery (SRS) and intensity modulated radiotherapy (IMRT). Precise measurement of dosimetric parameters such as beam profile, percentage depth doses and output factor of these beams are complicated due to the electron disequilibrium and the steep dose gradients. In the p...
متن کاملDosimetric comparison between XR-RV3 and EBT2 radiochromic film in megavoltage photon beams
Background: Radiochromic film is used for radiation dose measurement, XR-RV3 is used in fluoroscopy and EBT2 film in radiation therapy. The aim was to determine if the dosimetric properties of these two films are comparable in megavolt photon beams. Materials and Methods: Comparison measurements included: calibration curves, heterogeneous phantom dose profiles, and nasopharynx dose distributio...
متن کاملDose Calculations for Lung Inhomogeneity in High-Energy Photon Beams and Small Beamlets: A Comparison between XiO and TiGRT Treatment Planning Systems and MCNPX Monte Carlo Code
Introduction Radiotherapy with small fields is used widely in newly developed techniques. Additionally, dose calculation accuracy of treatment planning systems in small fields plays a crucial role in treatment outcome. In the present study, dose calculation accuracy of two commercial treatment planning systems was evaluated against Monte Carlo method. Materials and Methods Siemens Once or linea...
متن کامل