Phantom Validation of Tc-99m Absolute Quantification in a SPECT/CT Commercial Device

نویسندگان

  • Silvano Gnesin
  • Paulo Leite Ferreira
  • Jerome Malterre
  • Priscille Laub
  • John O. Prior
  • Francis R. Verdun
چکیده

Aim. Similar to PET, absolute quantitative imaging is becoming available in commercial SPECT/CT devices. This study's goal was to assess quantitative accuracy of activity recovery as a function of image reconstruction parameters and count statistics in a variety of phantoms. Materials and Methods. We performed quantitative 99mTc-SPECT/CT acquisitions (Siemens Symbia Intevo, Erlangen, Germany) of a uniform cylindrical, NEMA/IEC, and an anthropomorphic abdominal phantom. Background activity concentrations tested ranged: 2-80 kBq/mL. SPECT acquisitions used 120 projections (20 s/projection). Reconstructions were performed with the proprietary iterative conjugate gradient algorithm. NEMA phantom reconstructions were obtained as a function of the iteration number (range: 4-48). Recovery coefficients, hot contrast, relative lung error (NEMA phantom), and image noise were assessed. Results. In all cases, absolute activity and activity concentration were measured within 10% of the expected value. Recovery coefficients and hot contrast in hot inserts did not vary appreciably with count statistics. RC converged at 16 iterations for insert size > 22 mm. Relative lung errors were comparable to PET levels indicating the efficient integration of attenuation and scatter corrections with adequate detector modeling. Conclusions. The tested device provided accurate activity recovery within 10% of correct values; these performances are comparable to current generation PET/CT systems.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Phantom study of fusion image of CT and SPECT with body-contour generated from external Compton scatter sources.

PURPOSE A phantom study was conducted to evaluate the feasibility of body contour definition with Compton scatter photons from external sources of technetium-99m pertechnetate (Tc-99m) to create a fusion image of CT and SPECT images. METHODS External sources of 1 mCi (37 MBq) Tc-99m were placed on each collimator, and body-contour SPECT images were obtained with an energy window of 100 keV +/...

متن کامل

Direct comparison of 99mTc-PSMA SPECT/CT and 68Ga-PSMA PET/CT in patients with prostate cancer

Objective(s): 99mTc-PSMA SPECT/CT is a cost effective alternative for 68Ga-PSMA PET/CT. The aim of this study was to directly compare these two techniques in patients with prostate cancer. Methods: 28 man with prostate cancer were studied using 99mTc-PSMA SPECT/CT and 68Ga-PSMA PET/CT in a shor...

متن کامل

Multiple brown tumors in a patient with parathyroid adenoma depicting affinity for both 99mTc -pertechnetate and 99mTc-MIBI: Evaluation with hybrid SPECT/CT

 Parathyroid adenoma involving a single parathyroid gland is the underlying cause of 80-85% of primary hyperparathyroidism (PHPT). Skeletal system is significantly affected by PHPT. Brown tumors are known to have affinity for 99mTc-MIBI. We report a rare case of PHPT presenting with diffuse bony pain, high calcium level and significantly elevated alkaline phosphatase level. 99m<...

متن کامل

Organ-specific SPECT activity calibration using 3D printed phantoms for molecular radiotherapy dosimetry

BACKGROUND Patient-specific absorbed dose calculations for molecular radiotherapy require accurate activity quantification. This is commonly derived from Single-Photon Emission Computed Tomography (SPECT) imaging using a calibration factor relating detected counts to known activity in a phantom insert. METHODS A series of phantom inserts, based on the mathematical models underlying many clini...

متن کامل

Abdo-Man: a 3D-printed anthropomorphic phantom for validating quantitative SIRT

BACKGROUND The use of selective internal radiation therapy (SIRT) is rapidly increasing, and the need for quantification and dosimetry is becoming more widespread to facilitate treatment planning and verification. The aim of this project was to develop an anthropomorphic phantom that can be used as a validation tool for post-SIRT imaging and its application to dosimetry. METHOD The phantom de...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2016  شماره 

صفحات  -

تاریخ انتشار 2016