[Comparison of FBP and ML-EM reconstruction used by simulation data; effect of the projection number for the SPECT image].

نویسندگان

  • Shuuji Fujihara
  • Shingo Kageyama
  • Yasunori Isoda
  • Akio Nagaki
  • Norikazu Matsutomo
  • Akira Takahata
  • Yoshihiro Komi
  • Hideo Onishi
چکیده

PURPOSE The influence of the numbers of projection of SPECT exerting on a re-constructed image cannot be strictly evaluated by phantom studies. Therefore, we compared re-constructed images of the FBP method and the ML-EM method by using simulation data. METHODS Simulation data was entered in the image processing software, and the projection data that changed the numbers of projection was made. Afterwards, reconstructed images of the FBP and the ML-EM methods were compared with respect to contrast, %COV, and the NMSE value. RESULT When the numbers of projection of the FBP and the ML-EM method were decreased, all of the contrast, %COV, and the NMSE value were more deteriorated than that of the ideal image. Therefore, the image quality of SPECT improves with both FBP and ML-EM methods when there are many numbers of projection. Moreover, the FBP method was excellent in a cold contrast, and the ML-EM method was uniformly excellent. Therefore, an understanding of features and their inspection are effective for the selection of each image reconstruction method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Utilization of an optimum low-pass filter during filtered back-projection in the reconstruction of single photon emission computed tomography images of small structures

Introduction:Low-pass filters eliminate noise, and accordingly improve the quality of filtered back-projection (FBP) in the reconstruction of single photon emission computed tomography (SPECT) images. This study aimed at selection of an optimum low-pass filter for FBP reconstruction of SPECT images of small structures. Material and Methods:Sp...

متن کامل

Performance Evaluation of FBP Reconstruction in SPECT Imaging

Introduction:  The  purpose  of  this  study  is  to  define  the  optimal  parameters  for  the  tomographic  reconstruction procedure in a routine single photon emission tomography. The Hoffman brain phantom  is modified to evaluate the reconstruction method. The phantom was imaged in a 3 and 2-dimensional  conformation and the results were compared.   Materials  and  Methods:  The  2D  phant...

متن کامل

Assessment of simulated patient motion and its effect on myocardial perfusion SPECT using two reconstruction methods (Filtered Backprojection;FBP and Iterative method) [Persian]

Introduction: Motion of the patient during myocardial perfusion SPECT could potentially results in false perfusion defects. The effect of different reconstruction methods on these artifacts is not studied. Clarification of the relation between the extent, severity and duration of motion with the resultant artifacts may be helpful in designing special soft wares for motion correction. This...

متن کامل

Evaluation of iterative reconstruction method and attenuation correction on brain dopamine transporter SPECT using anthropomorphic striatal phantom

Objective(s): The aim of this study was to determine the optimal reconstruction parameters for iterative reconstruction in different devices and collimators for dopamine transporter (DaT) single-photon emission computed tomography (SPECT). The results were compared between filtered back projection (FBP) and different attenuation correction (AC) methods.Methods: An anthropomorphic striatal phant...

متن کامل

Determination of the optimum filter for qualitative and quantitative 99mTc myocardial SPECT imaging

Background: Butterworth, Gaussian, Hamming, Hanning, and Parzen are commonly used SPECT filters during filtered back-projection (FBP) reconstruction, which greatly affect the quality and size accuracy of image. Materials and Methods: This study involved a cardiac phantom in which 1.10 cm thick cold defect was inserted into its myocardium wall and filled with 4.0 μCi/ml (0.148 MBq/ml) 99mTc conc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Nihon Hoshasen Gijutsu Gakkai zasshi

دوره 66 12  شماره 

صفحات  -

تاریخ انتشار 2010