Intrathoracic current flow during transthoracic defibrillation in dogs. Transcardiac current fraction.
نویسندگان
چکیده
To achieve transcardiac threshold current during transthoracic defibrillation, a considerably larger current must be delivered to the thorax to compensate for the shunting effect of the lungs, the thoracic cage, and other elements of the torso. This shunting effect is thus an important determinant of transthoracic defibrillation threshold and can be quantified by the transcardiac current fraction (FC, the ratio of transcardiac to transthoracic threshold currents). Previous estimates of FC have ranged from as low as 3% to as high as 45%. The purpose of of this study was to quantify both FC and the major intrathoracic current pathways. Transthoracic and intrathoracic voltages and currents were simultaneously measured during high-voltage transthoracic shocks in 20 dogs. With correction factors determined from another set of 12 dogs, these raw data were corrected to compensate for field distortion caused by the presence of the intrathoracic electrodes, and the adjusted data were fit to a resistive network model. The results showed that 82% of the transthoracic current was shunted by the thoracic cage, while 14% was shunted by the lungs. The remaining 4% (FC) is the portion that passed through the heart. There was good agreement between the two independent methods used to calculate FC. Analysis based on the model indicated that FC was 3.7%, whereas FC determined by direct measurement with calibrated electrodes was 4.2%. Therefore, the results of this study, in contrast to earlier estimates of FC, show that defibrillation in dogs is achieved by only 4% of the total transthoracic current.
منابع مشابه
Relation between transcardiac and transthoracic current during defibrillation in humans.
Conceptually, transthoracic defibrillation threshold current can be considered a function of at least two quantities. It is directly proportional to the transcardiac threshold current and inversely proportional to the transcardiac current fraction (FC) or the ratio of transcardiac and transthoracic current. Although experimental and theoretical estimates of FC have been as high as 45%, previous...
متن کاملRise in defibrillation threshold after postoperative cardiac remodeling in a patient with severe Ebstein’s anomaly
Introduction Implantable cardioverter-defibrillators (ICDs) are considered standard of care for patients with life-threatening cardiac arrhythmias. In small children and patients with venous anatomy that prohibits placement of traditional transvenous leads, nontransvenous coil positions have been used and can be characterized as epicardial, pleural, subcutaneous, or a hybrid of any of the 3 pos...
متن کاملdefibrillation: experimental studies
In defibrillation, current flow depends on the energy selected and the transthoracic impedance. If transthoracic impedance is high, current flow may be inadequate to defibrillate. We developed a method by which high transthoracic impedance is automatically compensated for by an increase in operator-selected energy when impedance is high. Transthoracic impedance was predicted in advance of the f...
متن کاملTransthoracic Resistance in Human Defibrillation
Successful defibrillation depends on delivery of adequate electrical current to the heart; one of the major determinants of current flow is transthoracic resistance (TTR). To study the factors influencing ITR, we prospectively collected data from 44 patients undergoing emergency defibrillation. Shocks of94-450 J delivered energy were administered from specially calibrated Datascope defibrillato...
متن کاملEffect of cervical vagal nerve stimulation on defibrillation energy: a possible adjunct to efficient defibrillation.
The efficacy of electrical defibrillation is considered to be related to the autonomic status. In search of a possible adjunct to enhance the therapeutic performance of an implantable cardioverter-defibrillator. we investigated whether parasympathetic manipulation by cervical vagal nerve stimulation (VNS) increases defibrillation efficacy. The effects of VNS on transcardiac defibrillation thres...