OBJECTIVE Percutaneous transcatheter occlusion has benefited thousands of patients suffering from patent foramen ovale and atrial septal defect. However, no general agreement has been reached on the superiority among occluders. Thus, a meta-analysis between the two most commonly adopted types of occluders was conducted. METHODS The literature review has identified relevant studies up to May, 2011 in the databases of PubMed, EMBASE, the Cochrane Central Register of Controlled Trials, and World Health Organization clinical trials registry centre. Meta-analysis was performed in a fixed/random effects model using Revman 5.1.1. Information on complications and outcomes was extracted. RESULTS Analysis from included studies reports an outcome in favour of the Amplatzer. The Amplatzer has proven its superiority in efficacy with a significantly lower risk of early (95% confidence interval = 0.09-0.34) and long-term (95% confidence interval = 0.14-0.97) residual shunt rate for atrial septal defect occlusion, although no significant difference in performance has been reported for patent foramen ovale. In addition, the Amplatzer has also remarkably reduced the risk of embolisation by the device (95% confidence interval = 0.07-0.45) for atrial septal defect and new-set atrial fibrillation (95% confidence interval = 0.18-0.48) for patent foramen ovale. On evaluation of recurrent thrombotic events, it was found that the Amplatzer greatly lowered the rate of thrombus formation on the device (95% confidence interval = 0.02-0.21) for patent foramen ovale; however, no statistical difference was found on atrial septal defect evaluation. However, the result indicated no statistically significant difference between the two kinds of occluders in stroke and transient ischaemic attack of patent foramen ovale. CONCLUSION The meta-analysis has proven the Amplatzer to be the superior occluder, serving better prognosis with more fluent procedure and less complications.
منابع مشابه
Assessment of the geometric profile of the Amplatzer and Cardioseal septal occluders by three dimensional echocardiography.
OBJECTIVE To apply three dimensional echocardiography to describe the geometric profile of the Amplatzer and Cardioseal occluders after deployment for closure of atrial septal defect. METHODS 20 patients (mean (SD) age, 14 (5) years) were enrolled for transcatheter closure of a secundum atrial septal defect with the Amplatzer occluder (10) or with the Cardioseal occluder (10). The two populat...
متن کاملLong-term results of a randomized trial comparing three different devices for percutaneous closure of a patent foramen ovale.
AIMS Percutaneous patent foramen ovale (PFO) closure for secondary stroke prevention is discussed controversially. Long-term data comparing different closure devices are limited. The objective is the prospective comparison of procedural complications and long-term results after PFO closure in patients with cryptogenic stroke randomized to three different closure devices. METHODS AND RESULTS B...
متن کاملTranscatheter Amplatzer device closure of atrial septal defect and patent foramen ovale in patients with presumed paradoxical embolism.
OBJECTIVE To review our experience with, and profile the safety and efficacy of, the Amplatzer PFO (patent foramen ovale) occluder (APO) and Amplatzer septal occluder (ASO) used to close PFO and/or atrial septal defect (ASD) in patients with paradoxical embolism (PE). PATIENTS AND METHODS Between April 1998 and November 2002, 103 patients at the Mayo Clinic in Rochester, Minn, and Scottsdale,...
متن کاملEarly and late complications associated with transcatheter occlusion of secundum atrial septal defect.
OBJECTIVES The goal of this study was to report the early and late complications experienced in atrial septal defect (ASD) transcatheter closure. BACKGROUND Atrial septal defect transcatheter occlusion techniques have become an alternative to surgical procedures. A number of different devices are available for transcatheter ASD closure. The type and rate of complications are different for dif...
متن کاملTranscatheter closure of atrial septal defects without fluoroscopy: feasibility of a new method.
BACKGROUND In an effort to reduce x-ray exposure, we developed a technique for transcatheter closure of atrial septal defects under echocardiographic guidance without fluoroscopy. To assess the efficiency of this procedure for routine use, we compared our initial results with those for the conventional procedure. METHODS AND RESULTS Twenty-two randomly selected patients (median age 18 years; ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Cardiology in the young
دوره 23 4 شماره
صفحات -
تاریخ انتشار 2013