The accuracy of dynamic wedge dose computation in the ADAC Pinnacle RTP system

نویسندگان

  • H. Shao
  • X. Wu
  • C. Luo
  • S. Crooks
  • A. Bernstein
  • A. Markoe
چکیده

The nonphysical wedge is a modality that uses computer-controlled jaw motion to generate wedge-shaped dose distributions. There are Varian enhanced dynamic wedges (EDWs) and Siemens virtual wedges (VWs). We recently commissioned dynamic wedges on both Varian and Siemens LINACs. The beam data, acquired with a Wellhofer chamber array and a Sun Nuclear profiler, are used for modeling in the ADAC Pinnacle system. As recommended by ADAC, only a limited number of beam data is measured and used for beam modeling. Therefore, the dose distributions of dynamic wedges generated by Pinnacle must be examined. Following the commissioning of the dynamic wedges, we used Pinnacle to generate a number of dose distributions with different energies, wedge angles, field sizes, and depths. The computed data from Pinnacle are then compared with the measured data. The deviations of the output factor in all square and rectangular fields are mostly within 2.0% for both EDW and VW. For asymmetric fields, the deviations are within 3%. However, exceptions of differences more than 3% have been found in a larger field and large wedge combinations. The precision of the beam profiles generated by Pinnacle is also evaluated. As a result of this investigation, we present a scope of quality assurance tests that are necessary to ensure acceptable consistency between the delivered dose and the associated treatment plan when dynamic wedges are applied.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Beam‐commissioning methodology for a three‐dimensional convolution/superposition photon dose algorithm

Commissioning beam data for the convolution/superposition dose-calculation algorithm used in a commercial three-dimensional radiation treatment planning (3D RTP) system (PINNACLE(3), ADAC Laboratories, Milpitas, CA) can be difficult and time consuming. Sixteen adjustable parameters, as well as spectral weights representing a discrete energy spectrum, must be fit to sets of central-axis depth do...

متن کامل

Optimum reckoning of contra lateral breast dose using physical wedge and enhanced dynamic wedge in radiotherapy treatment planning system

Background: It was intended to investigate the effect of physical wedge (PW) and enhanced dynamic wedges (EDW) on contralateral breast dose during primary breast irradiation in radiotherapy treatment, using high energy photon beams. Materials and Methods: The Varian’s Clinac dual mode linear accelerator model 2100 C/D and Siemen’s Primus accelerators were used for radiation doses wi...

متن کامل

Assessment of Dose Calculation Accuracy of TiGRT Treatment Planning System for Physical Wedged fields in Radiotherapy

Introduction Wedge modifiers are commonly applied in external beam radiotherapy to change the dose distribution corresponding to the body contour and to obtain a uniform dose distribution within the target volume. Since the radiation dose delivered to the target must be within ±5% of the prescribed dose, accurate dose calculation by a treatment planning system (TPS) is important. The objective ...

متن کامل

Evaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry

Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...

متن کامل

Dosimetric evaluation of a treatment planning system using pencil beam convolution algorithm for enhanced dynamic wedges with symmetric and asymmetric fields

Background: The dosimetric performance of Eclipse 6.5 three dimensional treatment planning system (3DTPS) is evaluated by comparing the calculated and measured dose in two dimensions following the guide lines of American Association for Physicists in Medicine Task Group 53. Materials and Methods: The calculations were performed by the 3DTPS for symmetric as well as asymmetric fields for standar...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2004