Absorbed dose measurements of a handheld 50 kVP X-ray source in water with thermoluminescence dosemeters.
نویسندگان
چکیده
Absorbed dose rate measurements of a 50 kV(p) handheld X-ray probe source in a water phantom are described. The X-ray generator is capable of currents of up to 40 microA, and is designed for cranial brachytherapy and intraoperative applications with applicators. The measurements were performed in a computer-controlled water phantom in which both the source and the detectors are mounted. Two different LiF thermoluminescence dosemeter (TLD) phosphors were employed for the measurements, MTS-N (LiF:Mg,Ti) and MCP-N (LiF:Mg,Cu,P). Two small ionisation chambers (0.02 and 0.0053 cm(3)) were also employed. The TLDs and chambers were positioned in watertight mounts made of water-equivalent plastic. The chambers were calibrated in terms of air-kerma rate, and conventional protocols were used to convert the measurements to absorbed dose rate. The TLDs were calibrated at National Institute of Standards and Technology (NIST) in terms of absorbed dose rate using a (60)Co teletherapy beam and narrow-spectrum X-ray beams. For the latter, absorbed dose was inferred from air-kerma rate using calculated air-kerma-to-dose conversion factors. The reference points of the various detectors were taken as the center of the TLD volumes and the entrance windows of the ionisation chambers. Measurements were made at distances of 3-45 mm from the detector reference point to the source center. In addition, energy dependence of response measurements of the TLDs used was made using NIST reference narrow spectrum X-ray beams. Measurement results showed reasonable agreement in absorbed dose rate determined from the energy dependence corrected TLD readings and from the ionisation chambers. Volume averaging effects of the TLDs at very close distances to the source were also evident.
منابع مشابه
Monte Carlo radiography simulation for assessment of absorbed radiation dose in femur bone marrow during X-ray radiography for constant mAs and AEC techniques
Background: The purpose of this study was to find the accurate absorbed dose in the femur bone marrow during the X-ray radiography for constant mAs and AEC techniques. Materials and Methods: The DOSXYZnrc was used to simulate radiation doses in two human femurs during diagnostic radiography. EGSnrc phantoms produced from actual CT images of human femurs were modified by adding seven micrometre ...
متن کاملDosimetry of Occupational Radiation around Panoramic X-ray Apparatus
Background: Panoramic imaging is one of the most common imaging methods in dentistry. Regarding the side-effects of ionizing radiation, it is necessary to survey different aspects and details of panoramic imaging. In this study, we compared the absorbed x-ray dose around two panoramic x-ray units: PM 2002 CC Proline (Planmeca, Helsinki, Finland) and Cranex Tome (Soredex, Helsinki, Finland).<br ...
متن کاملEvaluation of depth dose characteristics of superficial X-rays machine using different kVp and applicators diameter
Background: Radiotherapy Treatment Planning requires different dosimetric quantities as input in order to calculate a desired dose distribution. This study has been focused to evaluate the depth dose characteristics of superficial X-rays being used for radiotherapy treatment. Materials and Methods: Computerized 3-D water phantom of multi-data system was used. The measurements were made...
متن کاملSynthesis and thermoluminescence properties investigation of nano-structured alumina for high dose measurement
In this study, nanostructured alumina was synthesized by solution combustion method and its thermoluminescence properties was investigated. X-raydiffraction (XRD) and scanning electron microscopy (SEM) were conducted to investigate the crystal structure, morphology and the particle size of the powders. It was observed that nano structured α-Al2O3 particles with nano flake morphology and thickne...
متن کاملA survey of annual effective and genetically significant dose from conventional X-ray examinations in 10 counties in Khorasan province-Iran
Background: A study has been conducted to assess two useful radiation protection indices, the annual per-capita effective dose equivalent (AED) and the genetically significant dose (GSD), from conventional diagnostic radiography in 10 counties with more than 1,886,000 inhabitants in Khorasan province in Iran. Materials and Methods: The health centre authorities of Khorasan province were asked t...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Radiation protection dosimetry
دوره 120 1-4 شماره
صفحات -
تاریخ انتشار 2006