Piezoresistive AFM cantilevers surpassing standard optical beam deflection in low noise topography imaging
نویسندگان
چکیده
Optical beam deflection (OBD) is the most prevalent method for measuring cantilever deflections in atomic force microscopy (AFM), mainly due to its excellent noise performance. In contrast, piezoresistive strain-sensing techniques provide benefits over OBD in readout size and the ability to image in light-sensitive or opaque environments, but traditionally have worse noise performance. Miniaturisation of cantilevers, however, brings much greater benefit to the noise performance of piezoresistive sensing than to OBD. In this paper, we show both theoretically and experimentally that by using small-sized piezoresistive cantilevers, the AFM imaging noise equal or lower than the OBD readout noise is feasible, at standard scanning speeds and power dissipation. We demonstrate that with both readouts we achieve a system noise of ≈0.3 Å at 20 kHz measurement bandwidth. Finally, we show that small-sized piezoresistive cantilevers are well suited for piezoresistive nanoscale imaging of biological and solid state samples in air.
منابع مشابه
Development of liquid-environment frequency modulation atomic force microscope with low noise deflection sensor for cantilevers of various dimensions
We have developed a liquid-environment frequency modulation atomic force microscope FM-AFM with a low noise deflection sensor for a wide range of cantilevers with different dimensions. A simple yet accurate equation describing the theoretical limit of the optical beam deflection method in air and liquid is presented. Based on the equation, we have designed a low noise deflection sensor. Replace...
متن کاملAdvances in self-sensing techniques for atomic force microscopy
Atomic force microscope (AFM) is a tool that allows micro and nano scale imaging of samples ranging from solid state physics to biology. AFM uses mechanical forces to sense the sample and recreate a topography image with high spatial resolution. The biggest disadvantage of the standard AFMs is their scanning speed, as it typically takes up to several tens of minutes to capture an image. A lot o...
متن کاملHigh-speed imaging upgrade for a standard sample scanning atomic force microscope using small cantilevers.
We present an atomic force microscope (AFM) head for optical beam deflection on small cantilevers. Our AFM head is designed to be small in size, easily integrated into a commercial AFM system, and has a modular architecture facilitating exchange of the optical and electronic assemblies. We present two different designs for both the optical beam deflection and the electronic readout systems, and...
متن کاملA scanning probe microscope for magnetoresistive cantilevers utilizing a nested scanner design for large-area scans
We describe an atomic force microscope (AFM) for the characterization of self-sensing tunneling magnetoresistive (TMR) cantilevers. Furthermore, we achieve a large scan-range with a nested scanner design of two independent piezo scanners: a small high resolution scanner with a scan range of 5 × 5 × 5 μm(3) is mounted on a large-area scanner with a scan range of 800 × 800 × 35 μm(3). In order to...
متن کاملWideband low-noise optical beam deflection sensor with photothermal excitation for liquid-environment atomic force microscopy.
I developed a wideband low-noise optical beam deflection sensor with a photothermal cantilever excitation system for liquid-environment atomic force microscopy. The developed sensor has a 10 MHz bandwidth and 4.7 fm/sq.rt.Hz deflection noise density in water. The theoretically limited noise performance (i.e., the noise level limited only by the photodiode shot noise) has been achieved in liquid...
متن کامل