Axial length measurement acquisition rates of two optical biometers in cataractous eyes
نویسنده
چکیده
PURPOSE To compare the ability of two optical biometers to acquire the axial length (AL) measurement in cataractous eyes. METHODS This prospective, comparative, single-center study comprised 105 eyes (63 patients). AL was acquired by the composite mean value of 20 measurements (composite-20 IM) and five measurements (composite-5 IM) (IOLMaster(®) 500 version 7.1 software), and the standard mean value of the first five measurements (standard-5 LS, Lenstar LS 900(®)). Anterior chamber depth (ACD) and average keratometry (K) readings were acquired. RESULTS AL was acquired in 83.8%, 92.4%, and 84.8% of eyes for the composite-5 IM, composite-20 IM, and standard-5 LS, respectively. Standard-5 LS AL measurements were significantly shorter (P<0.001). IOLMaster(®) 500-acquired ACD (corneal epithelium to lens) measurements were significantly shorter (P<0.001). IOLMaster(®) 500 average K measurements were significantly steeper (P<0.001). CONCLUSION The composite-20 IM had the highest AL acquisition success rate of the three versions evaluated. AL, ACD, and average Ks were statistically different between the two biometers, although the differences were clinically insignificant.
منابع مشابه
Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract
PURPOSE To evaluate axial length (AL) measurement failure rate with the IOLMaster (Carl Zeiss AG, Germany) and Lenstar LS 900 (Haag-Streit AG, Switzerland) in eyes with cataract. METHODS Two hundred and ninety-six eyes of 170 patients with cataract were enrolled. Cataract type and severity were graded using the Lens Opacities Classification System III (LOCS III) and AL measurements were attem...
متن کاملComparison of Repeatability and Agreement between Swept-Source Optical Biometry and Dual-Scheimpflug Topography
Purpose To assess the repeatability and agreement of parameters obtained with two biometers and to compare the predictability. Methods Biometry was performed on 101 eyes with cataract using the IOLMaster 700 and the Galilei G6. Three measurements were obtained per eye with each device, and repeatability was evaluated. The axial length (AL), anterior chamber depth (ACD), keratometry (K), white...
متن کاملComparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer.
PURPOSE To compare the repeatability and reproducibility of ocular biometry and intraocular lens (IOL) power obtained with a new optical biometer (AL-Scan) and a standard optical biometer (IOLMaster 500). SETTING Siriraj Hospital, Mahidol University, Bangkok, Thailand. DESIGN Prospective comparative study. METHODS Two independent operators measured eyes with cataract using both biometers....
متن کاملKeratometry in pediatric eyes with cataract.
OBJECTIVES To report the keratometry data of pediatric cataractous eyes (randomly selected single eye of bilateral cases; cataractous eye of unilateral cases) and to compare the keratometry data of the unilateral cataractous eye with the corresponding noncataractous fellow eye. METHODS Retrospective review of preoperative data of patients who had undergone cataract surgery before 18 years of ...
متن کاملImmersion A-scan compared with partial coherence interferometry: outcomes analysis.
PURPOSE To compare 2 methods of axial length measurement, immersion ultrasonography and partial coherence interferometry, and to elucidate surgical outcomes based on immersion measurements. SETTING Oregon Eye Institute, Eugene, Oregon, USA. METHODS Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound ...
متن کامل