'Fused-Gold' vs. 'Bare' stainless steel NIRflex stents of the same geometric design in diseased native coronary arteries. Long-term results from the NIR TOP Study.
نویسندگان
چکیده
OBJECTIVES We evaluated the long-term clinical and angiographic results of 'fused-gold' (NIRFlex Royal) and 'bare' (NIRFlex) stainless steel stents in patients undergoing percutaneous coronary intervention (PCI). BACKGROUND Recent studies have shown high clinical and angiographic restenosis rates following the intracoronary implantation of 'gold-coated' stainless steel stents. The new 'fused-gold' stent, with improved surface characteristics and flexibility, was developed to improve procedural and long-term results, while maintaining enhanced radiopacity. METHODS A total of 305 patients (358 lesions) with symptomatic native coronary artery disease (CAD) undergoing native vessel PCI were randomised to receive a 'fused-gold' (n=147) or 'bare' (n=158) stent. Primary endpoint was minimal luminal diameter (MLD) at 6 months angiographic follow-up. Secondary endpoints included technical and procedural success, major adverse cardiac events (MACE), target vessel failure (TVF), angiographic binary restenosis rates, and additional angiographic comparisons. RESULTS There were no major differences in the baseline angiographic variables or patient characteristics between the two groups, however there was a trend towards a higher risk in the 'fused-gold' stent group. Clinical and angiographic follow-up was 100% and 87% respectively. MLD at 6 months follow-up was smaller in the 'fused-gold' stent group compared to the 'bare' stent group (1.61+/-0.65 vs. 1.81+/-0.60 mm, respectively); Therefore, the null hypothesis of non-inferiority cannot be rejected (p=0.49); equivalency cannot be claimed for the two stent types. The 'fused-gold' stents were also associated with a higher angiographic binary and clinical restenosis rates (33 vs. 18%; p=0.002 & 26.9 vs. 20.3%; p<0.001, respectively). CONCLUSION The 'bare' NIRflex stent was associated with excellent long-term clinical and angiographic results. Taking into account the equivalence margin, the null hypothesis of non-equivalence between the 'fused-gold' NIRflex Royal stent and the 'bare' NIRflex stent cannot be rejected (p=0.49), so equivalence cannot be claimed for the two stent types.
منابع مشابه
Gold-coated NIR stents in porcine coronary arteries.
BACKGROUND As endovascular stents are altered to add functionality, eg, by adding radiopaque coatings, biocompatibility may suffer. METHODS AND RESULTS We examined the vascular response in porcine coronary arteries to stainless steel gold-coated NIR stents (7-cell, Medinol, Inc). Stents, 9 and 16 mm in length, were left bare or coated with a 7-microm layer of gold. Physical and material effec...
متن کاملA stent is not just a stent: Stent construction and design do matter in its clinical performance.
The undisputed superiority of stents over conventional balloon angioplasty has resulted in a plethora of stents in clinical use. Recent data, however, have indicated not all stent models are the same. Nuances in stent design and construction have impacted significantly on the immediate and long-term clinical outcome. Among the stainless steel stents, those with multicellular or tubular designs ...
متن کاملDevelopment of gold stents for the treatment of intracranial aneurysms: an experimental study in a canine model.
BACKGROUND AND PURPOSE Gold has often been used in medicine because of its radiopacity and flexibility. To perform stent-supported coil embolization of intracranial aneurysms, we prepared a gold stent and examined its flexibility, radiopacity, and thrombogenic properties in comparison with a stainless steel device implanted in vitro and in vivo. METHODS Gold stents were prepared by plating go...
متن کاملThe future of coronary stent design REVIEW aRTICLE
The methodology in interventional cardiology has historically evolved from diagnostic coronary angiography to balloon angioplasty, to the use of bare metal stents (BmSs), to their refinement to drug-eluting stents with a durable polymer, and is now on the verge of drug-eluting stents with further developed drug delivery approaches such as reservoir technology and the use of bioresorbable polyme...
متن کاملInfluence of metal alloy and the profile of coronary stents in patients with multivessel coronary disease
BACKGROUND In Brazil, despite the recommendations of the Brazilian Society of Hemodynamics and Interventional Cardiology, the National Health System has not yet approved the use of drug-eluting stents. In percutaneous coronary interventions performed in the public and part of the private health care system, bare metal stents are used as the only option. Therefore, new information on bare metal ...