The characteristics of Fuji IP Cassette Type PII and application for radiotherapy quality assurance tests and portal imaging

نویسندگان

  • H. S. Soh
  • N. M. Ung
  • K. H. Ng
چکیده

The advancement of digital imaging has prompted more medical institutions to go filmless. The computed radiography (CR) system is becoming an important tool not only in diagnostic imaging, but also in radiation oncology. A new CR system that was specially designed for the use in radiation oncology, Fuji IP cassette type PII has been introduced to the market in the middle of year 2006. This project aimed to study some basic physical characteristics of this new type of cassette and explore its application for performing quality assurance (QA) tests and portal imaging in radiotherapy. All the images were read by FCR 5000 Plus reader. The image was found to reach its saturation value of 1023 (due to the image was stored in 10 bits data) by depending on the sensitivity value being adjusted. The uniformity test gave the result of 0.12%. The cassette was used to perform the QA tests which are previously performed by using film. All the results met the specification as stated in AAPM Task Group 40. The comparison for the portal images of PortalVisionTM contrast-detail phantom showed that the spatial resolution of the images obtained by CR system (Fujifilm Co., Ltd., Tokyo, Japan) were better than the EPID (Varian Medical Systems, Inc., Palo Alto, USA) and film system (Eastman Kodak Co., New York, USA). The IP cassette type PII was found to be suitable as an alternative QA test tool and portal imaging in radiotherapy.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy

Introduction: Electronic portal imaging devices (EPIDs) provide two- and three-dimensional planar and volumetric cone beam images to improve the accuracy of radiation treatment delivery. Periodic quality assurance (QA) of EPIDs is essential for dosimetric verification in radiotherapy. In this study, a QA program was implemented to evaluate the function of the EPID to be confident in applying co...

متن کامل

A quality assurance program for an amorphous silicon electronic portal imaging device using in-house developed phantoms: a method development for dosimetry purposes

Background: Electronic portal imaging devices (EPIDs) play an important role in radiation therapy portal imaging, geometric and dosimetric verifications. A successful utilization of EPIDs for imaging and dosimetric purposes requires a reliable quality control process routine to be carried out regularly. In this study, two in-house phantoms were developed and analyzed for implementation in a qua...

متن کامل

Treatment couch positioning uncertainties using an EPID-based method

Introduction: The accuracy of patient positioning plays an important role for radiotherapy tasks. Short and automated Quality Assurance (QA) programs are becoming one of the challenging tasks in Radiotherapy. The current study, investigates the accuracy of treatment couch positioning with four degrees of freedom (4DoF) using a fast and accurate method based on an image acquired...

متن کامل

Pretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)

Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...

متن کامل

Comparative Study between Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) for Radiation Treatment Verifications

Introduction: Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) are the preferred tools of Image Guided Radiotherapy (IGRT) and Dose Guided Radiotherapy (DGRT) which have been used for Radiotherapy treatment verifications. As a result, the number of publications dealing with these two tools for radiation treatment verification has increased consi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2008