The fixation strength of six hamstring tendon graft fixation devices in anterior cruciate ligament reconstruction. Part II: tibial site.

نویسندگان

  • Petteri Kousa
  • Teppo L N Järvinen
  • Mika Vihavainen
  • Pekka Kannus
  • Markku Järvinen
چکیده

BACKGROUND Tibial fixation is more problematic than femoral fixation in anterior cruciate ligament reconstruction. HYPOTHESIS There is no difference in initial fixation strength among hamstring tendon graft tibial fixation devices. STUDY DESIGN Randomized experimental study. METHODS Each of six devices used to fix 120 quadrupled human semitendinosus-gracilis tendon grafts into porcine tibiae was tested 10 times with a single-cycle load-to-failure test and 10 times with a 1500-cycle loading test. Specimens surviving cyclic loading were subjected to a single-cycle load-to-failure test. RESULTS Intrafix (1332 N) was the strongest in the single-cycle load-to-failure test, followed by WasherLoc (975 N), tandem spiked washer (769 N), SmartScrew ACL (665 N), BioScrew (612 N), and SoftSilk (471 N). After cyclic-loading tests, Intrafix showed the lowest residual displacement (1.5 mm) and was also strongest (1309 N) in the single-cycle load-to-failure test after the cyclic-loading test, followed by WasherLoc (3.2 mm; 917 N). CONCLUSION The Intrafix provided clearly superior strength in the fixation of hamstring tendon grafts to the tibial drill hole. CLINICAL RELEVANCE Some caution may be warranted when using the implants that showed increased residual displacement, especially if aggressive rehabilitation is to be used. Preconditioning of the hamstring tendon graft-implant complex before tibial fixation is needed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The fixation strength of six hamstring tendon graft fixation devices in anterior cruciate ligament reconstruction. Part I: femoral site.

BACKGROUND Strength of graft fixation is the weakest link in anterior cruciate ligament reconstruction. HYPOTHESIS There is no difference in initial fixation strength between different hamstring tendon graft femoral fixation devices. STUDY DESIGN Randomized experimental study. METHODS Each of six devices was used in the fixation of 10 quadrupled human semitendinosus-gracilis tendon grafts...

متن کامل

Multistranded hamstring tendon graft fixation with a central four-quadrant or a standard tibial interference screw for anterior cruciate ligament reconstruction.

BACKGROUND Tibial fixation of hamstring tendon grafts has been the weak link in anterior cruciate ligament reconstruction. HYPOTHESIS Use of a central four-quadrant sleeve and screw provides superior fixation when compared with standard interference screw fixation. STUDY DESIGN Controlled laboratory study. METHODS In eight pairs of cadaveric knees each anterior cruciate ligament was recon...

متن کامل

Biomechanical Comparison Between Bashti Bone Plug Technique and Biodegradable Screw for Fixation of Grafts in Ligament surgery

  Background: Ligament reconstruction is a common procedure in orthopedic surgery. Although several popular techniques are currently in use, new methods are proposed for secure fixation of the tendon graft into the bone tunnel. Purposes: We sought to introduce our new technique of Bashti bone plug for fixation of soft tissue graft in anterior cruciate ligament (ACL) reconstruction and to compar...

متن کامل

Biomechanical comparison of hamstring and patellar tendon graft anterior cruciate ligament reconstruction techniques: The impact of fixation level and fixation method under cyclic loading.

PURPOSE To mechanically test different reconstruction techniques of the anterior cruciate ligament (ACL) under incremental cyclic loading and to evaluate the impact of the level and method of graft fixation on tensile properties of each technique. TYPE OF STUDY In vitro biomechanical study. METHODS Four hamstring and 1 patellar tendon reconstruction techniques were performed on 40 young to ...

متن کامل

Novel Femoral Fixation Of Quadrupled ACL Reconstruction Graft Using Cancellous Bracing. Biomechanical Comparison With Established Devices

INTRODUCTION. The anterior cruciate ligament (ACL) is the most commonly injured major ligament of the knee (1). There are over 300,000 annually reported ACL injuries, a majority of which are complete ruptures (2). Long term success rates of 69-90% have been reported after arthrosopically-assisted ACL reconstruction, with a majority of failures related to graft fixation and surgical error. The e...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The American journal of sports medicine

دوره 31 2  شماره 

صفحات  -

تاریخ انتشار 2003