Quantitative Perfusion Analysis of First-Pass Contrast Enhancement Kinetics: Application to MRI of Myocardial Perfusion in Coronary Artery Disease

نویسندگان

  • Sohae Chung
  • Binita Shah
  • Pippa Storey
  • Sohah Iqbal
  • James Slater
  • Leon Axel
چکیده

PURPOSE Perfusion analysis from first-pass contrast enhancement kinetics requires modeling tissue contrast exchange. This study presents a new approach for numerical implementation of the tissue homogeneity model, incorporating flexible distance steps along the capillary (NTHf). METHODS The proposed NTHf model considers contrast exchange in fluid packets flowing along the capillary, incorporating flexible distance steps, thus allowing more efficient and stable calculations of the transit of tracer through the tissue. We prospectively studied 8 patients (62 ± 13 years old) with suspected CAD, who underwent first-pass perfusion CMR imaging at rest and stress prior to angiography. Myocardial blood flow (MBF) and myocardial perfusion reserve index (MPRI) were estimated using both the NTHf and the conventional adiabatic approximation of the TH models. Coronary artery lesions detected at angiography were clinically assigned to one of three categories of stenosis severity ('insignificant', 'mild to moderate' and 'severe') and related to corresponding myocardial territories. RESULTS The mean MBF (ml/g/min) at rest/stress and MPRI were 0.80 ± 0.33/1.25 ± 0.45 and 1.68 ± 0.54 in the insignificant regions, 0.74 ± 0.21/1.09 ± 0.28 and 1.54 ± 0.46 in the mild to moderate regions, and 0.79 ± 0.28/0.63 ± 0.34 and 0.85 ± 0.48 in the severe regions, respectively. The correlation coefficients of MBFs at rest/stress and MPRI between the NTHf and AATH models were r = 0.97/0.93 and r = 0.91, respectively. CONCLUSIONS The proposed NTHf model allows efficient quantitative analysis of the transit of tracer through tissue, particularly at higher flow. Results of initial application to MRI of myocardial perfusion in CAD are encouraging.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Myocardial perfusion scan accuracy in detection of coronary artery disease - Comparison with exercise stress test [Persian]

Introduction: In patients with coronary artery disease (CAD) noninvasive evaluation for detection of ischemia is important to avoid invasive interventions like angiography. Exercise stress test is conventionally the first study used in evaluation of CAD. Considering the noninvasive nature of the myocardial perfusion scan, we decided to compare its accuracy with stress test. Methods: Patie...

متن کامل

A Parametric Model for Quantitative Analysis of Contrast-Enhanced First-Pass MR Myocardial Perfusion that Accounts for Gd-DTPA Interstitial Loading

Introduction: Quantitative analysis of contrast-enhanced first-pass MR myocardial perfusion is useful for the detection of coronary artery disease. Myocardial blood flow (MBF) estimated from both arterial and myocardial time signal intensity curves provides more linear dynamic range for the assessment of myocardial ischemia than using semi-quantitative indices such as upslope and contrast enhan...

متن کامل

The first performance of myocardial perfusion scintigraphy, post intravenous Dipyridamole injection

SUMMARY Dipyridamolc-Thallium imaging is a relatively safe and accurate method to evaluate myocardial perfusion in patients who arc unable to exercise to a maximal exertion (5,6,8,9). A 51year-old woman was admitted to Taleghani hospital with clinical symptoms and signs of coronary artery disease. Intravenous dipyridamote-Thallium scan was performed in this case. In multiple views obtained f...

متن کامل

Relationship between the location of the most severe myocardial perfusion defects, the most severe coronary artery stenosis and the site of subsequent myocardial infarction [Persian]

This study evaluated the relationship between the location of the most severe coronary artery stenosis and the subsequent acute myocardial infarction (AMI). Methods: Of 1590 patients who were admitted with a diagnosis of AMI, we identified 44 patients who had undergone previous myocardial perfusion SPECT. Thirty-one of these patients also had previous coronary angiography. The relationshi...

متن کامل

Filter selection for 99mTc-Sestamibi myocardial perfusion SPECT imaging [Persian]

Myocardial perfusion SPECT imaging with Tc99m-sestamibi is the most accurate non-invasive means of detecting coronary artery disease and assessing the severity of perfusion abnormalities in the patients with coronary stenosis. Though simple and straight forward the results produced by the technique is very much affected by the details being used. One of the main problems with the test, wh...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2016