IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119.

نویسندگان

  • Gary A Ezzell
  • Jay W Burmeister
  • Nesrin Dogan
  • Thomas J LoSasso
  • James G Mechalakos
  • Dimitris Mihailidis
  • Andrea Molineu
  • Jatinder R Palta
  • Chester R Ramsey
  • Bill J Salter
  • Jie Shi
  • Ping Xia
  • Ning J Yue
  • Ying Xiao
چکیده

AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...

متن کامل

Dosimetry tools and techniques for IMRT.

Intensity modulated radiation therapy (IMRT) poses a number of challenges for properly measuring commissioning data and quality assurance (QA) radiation dose distributions. This report provides a comprehensive overview of how dosimeters, phantoms, and dose distribution analysis techniques should be used to support the commissioning and quality assurance requirements of an IMRT program. The prop...

متن کامل

Dosimetric Validation of Volumetric Modulated Arc Therapy (VMAT) Using AAPM TG-119 Benchmark Plans in an Upgraded CLINAC 2100CD for Flattening Filter Free (FFF) Photon Beams

Background: Recently we have upgraded our Varian Clinac 2100CD with a 6MV FFF beam, this upgrade being the first of its kind in our country. Even though the dosimetric characteristics of FFF beams have been reported both in experimental and Monte Carlo studies, application in planning and delivery is complex. The aim of this study was to validate the commissioning of upgraded FFF beams dosimetr...

متن کامل

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

The purpose of this work was to perform a comprehensive comparison of static field intensity modulated radiation therapy (IMRT) for patient specific quality assurance (QA) in order to investigate commissioning of a XiO treatment planning system (TPS). QA measurements were evaluated for three of four test geometries provided in American Association of Physicists in Medicine (AAPM) Task Group Rep...

متن کامل

Estimation of local confidence limit for 6 MV photon beam IMRT system using AAPM TG 119 test protocol

1Department of Radiation Oncology, SevenHills Hospital, Mumbai, India 2Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai, India Received October 17, 2015; Revised January 02, 2016; Accepted January 16, 2016; Published Online February 03, 2016 Original Article Abstract Purpose: The aim of this study was to estimate local confidence limit for 6 MV photon beam based...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 36 11  شماره 

صفحات  -

تاریخ انتشار 2009