The performance of Dräger Oxylog ventilators at simulated altitude.
نویسندگان
چکیده
Ventilated patients frequently require transport by air in a hypobaric environment. Previous studies have demonstrated significant changes in the performance of ventilators with changes in cabin pressure (altitude) but no studies have been published on the function of modem ventilators at altitude. This experiment set out to evaluate ventilatory parameters (tidal volume and respiratory rate) of three commonly used transport ventilators (the Dräger Oxylog 1000, 2000 and 3000) in a simulated hypobaric environment. Ventilators were assessed using either air-mix (60% oxygen) or 100% oxygen and tested against models simulating a normal lung, a low compliance (Acute Respiratory Distress Syndrome) lung and a high-resistance (asthma) lung. Ventilators were tested at a range of simulated altitudes between sea level and 3048 m. Over this range, tidal volume delivered by the Oxylog 1000 increased by 68% and respiratory rate decreased by 28%. Tidal volume delivered by the Oxylog 2000 ventilator increased by 29% over the same range of altitudes but there was no significant change in respiratory rate. Tidal volume and respiratory rate remained constant with the Oxylog 3000 over the same range of altitudes. Changes were consistent with each ventilator regardless of oxygen content or lung model. It is important that clinicians involved in critical care transport in a hypobaric environment are aware that individual ventilators perform differently at altitude and that they are aware of the characteristics of the particular ventilator that they are using.
منابع مشابه
Inspiratory oscillatory flow with a portable ventilator: a bench study
INTRODUCTION We observed an oscillatory flow while ventilating critically ill patients with the Dräger Oxylog 3000 transport ventilator during interhospital transfer. The phenomenon occurred in paediatric patients or in adult patients with severe airway obstruction ventilated in the pressure-regulated or pressure-controlled mode. As this had not been described previously, we conducted a bench s...
متن کاملEvaluation of the ventilator-user interface of 2 new advanced compact transport ventilators.
BACKGROUND Mechanical ventilation during patient transport frequently utilizes compact portable pneumatic ventilators that have limited ventilator-settings options. New advanced transport ventilators should yield quality improvements, but their user-friendliness needs to be tested. OBJECTIVE To evaluate the ventilator-user interface of 2 new transport ventilators. METHODS This was a 2-cente...
متن کاملEvaluation of transport ventilators at mild simulated altitude: a bench study in a hypobaric chamber.
BACKGROUND Previous studies on ventilators used for air transport showed significant effects of altitude, in particular with regard to accuracy of the tidal volume (VT) and breathing frequency. The aim of the study was to evaluate transport ventilators under hypobaric conditions. METHODS We conducted a bench study of 6 transport ventilators in a Comex hypobaric chamber to simulate mild altitu...
متن کاملPerformance of portable ventilators at altitude.
BACKGROUND Aeromedical transport of critically ill patients requires continued, accurate performance of equipment at altitude. Changes in barometric pressure can affect the performance of mechanical ventilators calibrated for operation at sea level. Deploying ventilators that can maintain a consistent tidal volume (VT) delivery at various altitudes is imperative for lung protection when transpo...
متن کاملA comparison of leak compensation during pediatric noninvasive ventilation: a lung model study.
BACKGROUND Ventilators used for noninvasive ventilation (NIV) must be able to synchronize in the presence of system leaks. We compared the ability of 7 ICU ventilators and 3 dedicated NIV ventilators to compensate for leaks during pediatric NIV. METHODS Using a lung simulator, we compared the Maquet Servo-i, Dräger V500, Dräger Carina, Covidien PB840, Respironics V60, Respironics Vision, GE H...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Anaesthesia and intensive care
دوره 36 4 شماره
صفحات -
تاریخ انتشار 2008