Reliability of Corneal Dynamic Scheimpflug Analyser Measurements in Virgin and Post-PRK Eyes

نویسندگان

  • Xiangjun Chen
  • Aleksandar Stojanovic
  • Yanjun Hua
  • Jon Roger Eidet
  • Di Hu
  • Jingting Wang
  • Tor Paaske Utheim
  • Sanjoy Bhattacharya
چکیده

PURPOSE To determine the measurement reliability of CorVis ST, a dynamic Scheimpflug analyser, in virgin and post-photorefractive keratectomy (PRK) eyes and compare the results between these two groups. METHODS Forty virgin eyes and 42 post-PRK eyes underwent CorVis ST measurements performed by two technicians. Repeatability was evaluated by comparing three consecutive measurements by technician A. Reproducibility was determined by comparing the first measurement by technician A with one performed by technician B. Intraobserver and interobserver intraclass correlation coefficients (ICCs) were calculated. Univariate analysis of covariance (ANCOVA) was used to compare measured parameters between virgin and post-PRK eyes. RESULTS The intraocular pressure (IOP), central corneal thickness (CCT) and 1st applanation time demonstrated good intraobserver repeatability and interobserver reproducibility (ICC ≧ 0.90) in virgin and post-PRK eyes. The deformation amplitude showed a good or close to good repeatability and reproducibility in both groups (ICC ≧ 0.88). The CCT correlated positively with 1st applanation time (r = 0.437 and 0.483, respectively, p<0.05) and negatively with deformation amplitude (r = -0.384 and -0.375, respectively, p<0.05) in both groups. Compared to post-PRK eyes, virgin eyes showed longer 1st applanation time (7.29 ± 0.21 vs. 6.96 ± 0.17 ms, p<0.05) and lower deformation amplitude (1.06 ± 0.07 vs. 1.17 ± 0.08 mm, p < 0.05). CONCLUSIONS CorVis ST demonstrated reliable measurements for CCT, IOP, and 1st applanation time, as well as relatively reliable measurement for deformation amplitude in both virgin and post-PRK eyes. There were differences in 1st applanation time and deformation amplitude between virgin and post-PRK eyes, which may reflect corneal biomechanical changes occurring after the surgery in the latter.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of Scheimpflug Holladay equivalent keratometry readings after corneal refractive surgery.

PURPOSE To determine the accuracy of Pentacam Scheimpflug system Holladay equivalent keratometry (K) readings (EKRs) in calculating intraocular lens (IOL) power after corneal refractive surgery, including laser in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and radial keratotomy (RK). SETTING Jules Stein Eye Institute, David Geffen School of Medicine at University of Calif...

متن کامل

مقایسه‌ی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبه‌ی لنز داخل چشمی افراد با قرنیه‌ی طبیعی

Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...

متن کامل

Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik

Swept-source optical coherence tomography (OCT) is the latest advancement in anterior segment imaging. There are limited data regarding its performance after laser in situ keratomileusis (LASIK). We compared the reliability of swept-source OCT and Scheimpflug imaging for evaluation of corneal parameters in refractive surgery candidates with myopia or myopic astigmatism. Three consecutive measur...

متن کامل

Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy

PURPOSE To determine inter-device agreement for central corneal thickness (CCT) measurement among ultrasound pachymetry, rotating Scheimpflug imaging (Pentacam, Oculus, Wetzlar, Germany), and scanning slit corneal topography (Orbscan II, Bausch & Lomb, Rochester, NY, USA) in highly myopic eyes before and after photorefractive keratectomy (PRK). METHODS This prospective comparative study inclu...

متن کامل

Changes in posterior corneal elevations after combined transepithelial photorefractive keratectomy and accelerated corneal collagen cross-linking: retrospective, comparative observational case series

BACKGROUND To compare the changes in anterior and posterior corneal elevations after combined transepithelial photorefractive keratectomy (PRK) and accelerated corneal collagen cross-linking (CXL) and after PRK. METHODS Medical records of 82 eyes of 44 patients undergoing either combined transepithelial PRK and CXL (PRK-CXL group) or transepithelial PRK (PRK group) were examined retrospective...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 9  شماره 

صفحات  -

تاریخ انتشار 2014