Evaluation of Frova, single-use intubation introducer, in a manikin. Comparison with Eschmann multiple-use introducer and Portex single-use introducer.
نویسندگان
چکیده
In a randomised cross-over study, 48 anaesthetists attempted to place a Frova single-use introducer, an Eschmann multiple-use introducer and a Portex single-use introducer in the trachea of a manikin set up to simulate a grade 3 laryngoscopic view. The anaesthetists were blinded to success (tracheal placement) or failure (oesophageal placement). Successful placement (proportion, 95% confidence interval) of either the Frova introducer (65%, 50-77%) or the Eschmann introducer (60%, 46-73%) was significantly more likely than with the Portex introducer (8%, 3-20%). There were no significant differences between the success rates for the Frova and the Eschmann introducers. A separate experiment revealed that the peak force exerted by the Frova and Portex introducers was two to three times greater than that which could be exerted by the Eschmann introducer, p < 0.0001, indicating that the single-use introducers are more likely to cause tissue trauma during placement.
منابع مشابه
Airway injury caused by a Portex single-use bougie.
To the Editor: The tracheal tube introducer (bougie) is widely used to facilitate difficult intubation. However, single-use introducers, compared to reusable ones, appear to be more traumatic [1-3]. A 61-year-old woman with esophageal reflux and no predictive signs of difficult intubation received rapidsequence induction with application of cricoid pressure. Direct laryngoscopy revealed a Corma...
متن کاملCombined use of a McGrath® MAC video laryngoscope and Frova Intubating Introducer in a patient with Pierre Robin syndrome: a case report
Patients with Pierre Robin syndrome are characterized by micrognathia, retrognathia, glossoptosis, and respiratory obstruction and are prone to have a difficult-to-intubate airway. The McGrath® MAC video laryngoscope provides a better view of the glottis than a Macintosh laryngoscope, but it is not easy to insert an endotracheal tube through the vocal cords because a video laryngoscope has a mu...
متن کاملBougies as an aid for endotracheal intubation with the Airway Scope: bench and manikin comparison studies
BACKGROUND When encountering a difficult airway with an Airway Scope (AWS) a bougie can be inserted into the endotracheal tube in the AWS channel. The angulated tip of the bougie can be guided toward the glottis by rotating it. We tested the ease of rotating bougies (Venn reusable, Boussignac, Portex single-use, and Frova) in an endotracheal tube when placed in the AWS channel. METHODS Bench ...
متن کاملGlideScope and Frova Introducer for Difficult Airway Management
The introduction into clinical practice of new tools for intubation as videolaringoscopia has dramatically improved the success rate of intubation and the work of anesthesiologists in what is considered the most delicate maneuver. Nevertheless intubation difficulties may also be encountered with good anatomical visualization of glottic structures in videolaringoscopia. To overcome the obstacles...
متن کاملEschmann Introducer Through Laryngeal Mask Airway: A Cadaveric Trial of An Alternate Means of Rescue Intubation
STUDY OBJECTIVE Laryngeal mask airways (LMAs) are often used as airway rescue devices where laryngoscopy is difficult. The LMA does not protect the airway and is preferably replaced with a cuffed endotracheal tube. There are reports of cases where an Eschmann tracheal tube introducer (ETTI) was successfully used to bridge between a standard LMA and an endotracheal tube. This project was designe...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Anaesthesia
دوره 59 8 شماره
صفحات -
تاریخ انتشار 2004