A device for precision positioning and alignment of room lasers to diminish their contribution to patient setup errors

نویسندگان

  • Ivan A. Brezovich
  • Stephen Jordan
چکیده

This report presents an analysis of patient setup errors resulting from inaccurately positioned wall lasers. It suggests that laser beams should agree within 0.2 degree or better with the machine axes that they are delineating. For typical simulator and treatment rooms having wall-to-isocenter distances of 3 m, this requirement is satisfied when the beam-emitting aperture is mounted within about 1.0 cm from the intersection of the respective machine axis with the wall. To achieve the required precision, we developed and clinically tested a simple, inexpensive tool, the Laser Placer (LP). The essential component of the LP is a cube with mirror surfaces that is aligned with the machine axes using built-in spirit levels and the light field and cross hairs of the collimator. Wall, ceiling, and sagittal lasers are installed and aligned according to reflections of their beams by the cube, and reference lines provided by the LP. Measurements showed that, even in new accelerator installations performed by highly experienced technicians, wall lasers are often mounted off target by more than 1.5 cm. Such inaccuracies can contribute systematic errors of 2 mm or more to the random setup errors attributable to interfraction movement in patient anatomy. To keep setup errors to a minimum, medical physicists should check beam orthogonality in addition to beam congruence at isocenter as recommended by the TG-40 quality assurance protocol from the American Association of Physicists in Medicine.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Utilization of Electronic Portal Imaging Device (EPID) For Setup Verification and Determination of Setup Margin in Head and Neck Radiation Therapy

Introduction: Radiation therapy involves a multistep procedure; therefore, the error in patient set up is an inherent part of the treatment. Main purpose of this study was to determine the clinical target volume (CTV) to planning target volume (PTV) in head and neck cancer patients. Material and Methods: A total of 15 patients who had daily p...

متن کامل

Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System

Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. Material and Methods: ...

متن کامل

A Simulation Study on Patient Setup Errors in External Beam Radiotherapy Using an Anthropomorphic 4D Phantom

Introduction Patient set-up optimization is required in radiotherapy to fill the accuracy gap between personalized treatment planning and uncertainties in the irradiation set-up. In this study, we aimed to develop a new method based on neural network to estimate patient geometrical setup using 4-dimensional (4D) XCAT anthropomorphic phantom. Materials and Methods To access 4D modeling of motion...

متن کامل

Accuracy of the dose delivery in prostate cancer patients-using an electronic portal imaging device (EPID)

Background: To correct patient positioning errors (setup errors) during prostate cancer treatment using EPID and fiducial gold markers, to improve the accuracy of the dose delivery in these patients. Materials and Methods: Fifteen patients with localized prostate carcinoma after implantation of fiducial gold markers in their prostate gland underwent the five-field IMRT planning technique. The p...

متن کامل

Calculation of Inter- and Intra-Fraction Motion Errors at External Radiotherapy Using a Markerless Strategy Based on Image Registration Combined with Correlation Model

Introduction: A new method based on image registration technique and an intelligent correlation model to calculate. The present study aimed to propose inter- and intra-fraction motion errors in order to address the limitations of conventional Patient positioning methods. Material and Methods: The configuration of the markerless method wa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2007