The clinical effectiveness of the streamlined liner of pharyngeal airway (SLIPA™) compared with the laryngeal mask airway ProSeal™ during general anesthesia

نویسندگان

  • Yun Mi Choi
  • Su Man Cha
  • Hyun Kang
  • Chong Wha Baek
  • Yong Hun Jung
  • Young Cheol Woo
  • Jin Yun Kim
  • Gill Hoi Koo
  • Sun Gyoo Park
چکیده

BACKGROUND The aim of this study was to compare the streamlined liner of the pharynx airway (SLIPA), a new supraglottic airway device (SGA), with the laryngeal mask airway ProSeal (PLMA) during general anesthesia. METHODS Sixty patients were randomly allocated to two groups; a PLMA group (n = 30) or a SLIPA group (n = 30). Ease of use, first insertion success rate, hemodynamic responses to insertion, ventilatory efficiency and positioning confirmed by fiberoptic bronchoscopy were assessed. Lung mechanics data were collected with side stream spirometry at 10 minutes after insertion. We also compared the incidence of blood stain, incidence and severity of postoperative sore throat and other complications. RESULTS First attempt success rates were 93.3% and 73.3%, and mean insertion time was 7.3 sec and 10.5 sec in PLMA and SLIPA. There was a significant rise in all of hemodynamic response from the pre-insertion value at one minute following insertion of SLIPA. But, insertion of PLMA was no significant rise in hemodynamic response. There was no statistically significant difference in the mean maximum sealing pressure, gas leakage, lung mechanics data, gastric distension, postoperative sore throat and other complication between the two groups. Blood stain were noted on the surface of the device in 40% (n = 12) in the SLIPA vs. 6.7% (n = 2) in the PLMA. CONCLUSIONS The SLIPA is a useful alternative to the PLMA and have comparable efficacy and complication rates. If we acquire the skill to use, SLIPA may be considered as primary SGA devices during surgery under general anesthesia.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Brief review: the Cobra Perilaryngeal Airway (CobraPLA and the Streamlined Liner of Pharyngeal Airway (SLIPA) supraglottic airways.

PURPOSE To summarize the published literature related to two, new supraglottic airways (SGAs): the Cobra Perilaryngeal Airway (CobraPLA) and the Streamlined Liner of Pharyngeal Airway (SLIPA). SOURCE We identified original articles, through searches on Medline and PubMed, using the keywords 'Cobra', 'CobraPLA', 'Cobra-PLA', 'PLA', and 'SLIPA'. PRINCIPAL FINDINGS Searches identified 28 publi...

متن کامل

Comparison of the clinical effectiveness between the streamlined liner of pharyngeal airway (SLIPA) and the laryngeal mask airway by novice personnel

BACKGROUND The aim of this study was to compare the streamlined liner of the pharynx airway (SLIPA) with the classic laryngeal mask airway when used by novice personnel. METHODS There were 114 patients enrolled into this study who underwent general anesthesia were randomly allocated into one of 2 groups; LMA group (n = 57) or SLIPA group (n = 57). After insertion, insertion success rate, inse...

متن کامل

Limitation of Laryngeal Mask Airway Usage in Open-Neck Procedure

I read the manuscript entitled “Case Report of An Open Neck Procedure Complication Associated with Laryngeal Mask Airway Use” by Subramaniam T et al. This is a valuable manuscript describing iatrogenic complications during level-2 cervical lymph node biopsy, in which the airway is managed with laryngeal mask airway (LMA). During the procedure, the surgeon realized that the mass was actually the...

متن کامل

Comparing the efficacy and safety of laryngeal mask airway, streamlined liner of the pharyngeal airway and I-gel following tracheal extubation

Adverse events following surgical operations are common complications due to removal of tracheal tube in contrast to the tracheal intubation. Awareness about the new methods and strategies for tracheal tube extubation is necessary for a safe and successful extubation. Therefore, we aimed to assess the safety and efficacy of laryngeal mask airway (LMA), streamlined liner of the pharyngeal airway...