Continuous positive airway pressure in new-generation mechanical ventilators: a lung model study.
نویسندگان
چکیده
BACKGROUND A number of new microprocessor-controlled mechanical ventilators have become available over the last few years. However, the ability of these ventilators to provide continuous positive airway pressure without imposing or performing work has never been evaluated. METHODS In a spontaneously breathing lung model, the authors evaluated the Bear 1000, Drager Evita 4, Hamilton Galileo, Nellcor-Puritan-Bennett 740 and 840, Siemens Servo 300A, and Bird Products Tbird AVS at 10 cm H(2)O continuous positive airway pressure. Lung model compliance was 50 ml/cm H(2)O with a resistance of 8.2 cm H(2)O x l(-1) x s(-1), and inspiratory time was set at 1.0 s with peak inspiratory flows of 40, 60, and 80 l/min. In ventilators with both pressure and flow triggering, the response of each was evaluated. RESULTS With all ventilators, peak inspiratory flow, lung model tidal volume, and range of pressure change (below baseline to above baseline) increased as peak flow increased. Inspiratory trigger delay time, inspiratory cycle delay time, expiratory pressure time product, and total area of pressure change were not affected by peak flow, whereas pressure change to trigger inspiration, inspiratory pressure time product, and trigger pressure time product were affected by peak flow on some ventilators. There were significant differences among ventilators on all variables evaluated, but there was little difference between pressure and flow triggering in most variables on individual ventilators except for pressure to trigger. Pressure to trigger was 3.74 +/- 1.89 cm H(2)O (mean +/- SD) in flow triggering and 4.48 +/- 1.67 cm H(2)O in pressure triggering (P < 0.01) across all ventilators. CONCLUSIONS Most ventilators evaluated only imposed a small effort to trigger, but most also provided low-level pressure support and imposed an expiratory workload. Pressure triggering during continuous positive airway pressure does require a slightly greater pressure than flow triggering.
منابع مشابه
Accuracy of automatic tube compensation in new-generation mechanical ventilators.
OBJECTIVE To compare performance of flow-adapted compensation of endotracheal tube resistance (automatic tube compensation, ATC) between the original ATC system and ATC systems incorporated in commercially available ventilators. DESIGN Bench study. SETTING University research laboratory. SUBJECTS The original ATC system, Dräger Evita 2 prototype, Dräger Evita 4, Puritan-Bennett 840. INT...
متن کاملThe impact of closed endotracheal suctioning systems on mechanical ventilator performance.
BACKGROUND Closed endotracheal suctioning during mechanical ventilation is increasingly used, but its impact on ventilator function has not been fully studied. METHODS We evaluated the impact of closed suctioning with 11 critical-care ventilators, during assisted ventilation in pressure-support mode, pressure-assist/control mode, volume-assist/control mode, and during continuous positive airw...
متن کاملComparison of the Effects of Prone and Supine Positions on Abdominal Distention in the Premature Infants Receiving Nasal Continuous Positive Airway Pressure (NCPAP)
Background: Premature infants with respiratory distress syndrome (RDS) are in dire need of respiratory support with a ventilator. However, the high tidal volume of mechanical ventilation may cause lung injury, and researchers have been concerned with the use of nasal continuous positive airway pressure (NCPAP). NCPAP has concomitant side effects, such as abdominal distention, which might disrup...
متن کاملSudden Complete Versus Gradual Weaning from Nasal Continuous Positive Airway Pressure in Preterm Neonates: A Randomized Controlled Trial
Background: Continuous positive airway pressure (CPAP) is used as respiratory support in preterm neonates; however, the best weaning method has not yet been determined. In this study, we compared sudden complete and gradual weaning from nasal CPAP (NCPAP) in preterm newborns. Methods: This randomized controlled trial was conducted on 62 prete...
متن کاملComparison of Nasal Continuous Positive Airway Pressure Therapy with and without Prophylactic Surfactant in Preterm Neonates
Background: Preterm labor is at risk for respiratory distress syndrome (RDS) and sometimes requires to mechanical ventilation (MV) and surfactant therapy. Continuous positive airway pressure (CPAP) and Nasal Continuous positive airway pressure (NCPAP) are the methods of respiratory support especially for using in RDS of neonates. In other method surfactant is administered to babies via tracheal...