An in‐house protocol for improved flood field calibration of TrueBeam FFF cine imaging

نویسندگان

  • Austin M. Faught
  • Fang‐Fang Yin
  • Justus Adamson
چکیده

PURPOSE TrueBeams equipped with the 40 × 30 cm2 Electronic Portal Imaging Devices (EPIDs) are prone to image saturation at the image center when used with flattening filter free (FFF) photon energies. While cine imaging during treatment may not saturate because the beam is attenuated by the patient, the flood field calibration is affected when the standard calibration procedure is followed. Here, we describe the hardware and protocol to achieve improved image quality for this model of TrueBeam EPID. MATERIALS & METHODS A stainless steel filter of uniform thickness was designed to have sufficient attenuation to avoid panel saturation. The cine imaging flood field calibration was acquired with the filter in place for the FFF energies under the standard calibration geometry (SID = 150 cm). Image quality during MV cine was assessed with & without the modified flood field calibration using a low contrast resolution phantom and an anthropomorphic phantom. RESULTS When the flood field is acquired without the filter in place, a pixel gain artifact is clearly present in the image center which may be mis-attributed to panel saturation in the subject image. At the image center, the artifact obscured all low contrast inserts and was also visible on the anthropomorphic phantom. Using the filter for flood field calibration eliminates the artifact. CONCLUSION TrueBeams equipped with the 40 × 30 cm2 IDU can utilize a modified flood field calibration procedure for FFF photon energies that improves image quality for cine MV imaging.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

SU-E-T-479: Skin Dose from Flattening Filter Free Beams: A Monte Carlo Investigation.

PURPOSE Flattening filter free (FFF) beams in radiotherapy have advantages such as shorter treatment delivery time and lower out-of-field dose compared with conventional flattened beams. This study investigates in detail the skin dose induced by FFF beams from a TrueBeam accelerator (Varian Medical Systems, Palo Alto, CA) using Monte Carlo method. METHODS Phase space files generated using rea...

متن کامل

Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx

Small field dosimetry is a challenging task. The difficulties of small field measurements, particularly stereotactic field size measurements, are highlighted by the large interinstitution variability that can be observed for circular cone collimator commissioning measurements. We believe the best way to improve the consistency of small field measurements is to clearly document and share the res...

متن کامل

Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans

A Monte Carlo (MC) validation of the vendor-supplied Varian TrueBeam 6 MV flattened (6X) phase-space file and the first implementation of the Siebers-Keall MC MLC model as applied to the HD120 MLC (for 6X flat and 6X flattening filter-free (6X FFF) beams) are described. The MC model is validated in the context of VMAT patient-specific quality assurance. The Monte Carlo commissioning process inv...

متن کامل

 A novel approach to automatic position calibration for pixelated crystals in gamma imaging

Introduction: The position estimation in gamma detection system will have constant misplacements which can be corrected in the calibration procedure. In the pixelated crystal uniformly irradiation of detector will produce irregular shape due to position estimation errors. This image is called flood field image and is used to calibrate the position estimation. In this work we pr...

متن کامل

Ion recombination correction factors (Pion) for Varian TrueBeam high‐dose‐rate therapy beams

Ion recombination is approximately corrected for in the Task Group 51 protocol by Pion, which is calculated by a two-voltage measurement. This measurement approach may be a poor estimate of the true recombination, particularly if Pion is large (greater than 1.05). Concern exists that Pion in high-dose-per-pulse beams, such as flattening filter free (FFF) beams, may be unacceptably high, renderi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017