Prospective two-year subsidence analysis of 100 cemented polished straight stems - a short-term clinical and radiological observation
نویسندگان
چکیده
BACKGROUND Cemented stems show good long-term results and the survival of new implants can be predicted by their early subsidence. With EBRA-FCA (Femoral Component Analysis using Einzel-Bild-Röntgen-Analyse) early subsidence as an early indicator for later aseptic loosening can be analysed. For the cemented TwinSys stem mid- and long-term data is only avalible from the New Zeeland Arthroplasty register, thus close monitoring of this implant system is still mandatory. METHODS We conducted a 2 year follow up of 100 consecutive hybrid THA (Total hip arthroplasty) of a series of 285 primary THA operated between Jan 2009 und Oct 2010. These 100 received a polished, cemented collarless straight stem (twinSys®, Mathys AG® Bettlach, Switzerland) with an uncemented monobloc pressfit cup (RM pressfit®, Mathys AG® Bettlach, Switzerland). The other patients were treated with the uncemented version of this stem and the same cup. Clinical (Harris Hip Score) and radiological (ap and axial x-rays, cementing quality according to Barrack, alignment) outcomes besides an EBRA-FCA subsidence analysis were performed. RESULTS Median age at operation was 78 (68 to 93) years. 5 patients died in the course of follow-up unrelated to surgery. The KM (Kaplan-Meier) survival at 2 years for the endpoint reoperation for any reason was 94.9 (95 % confidence interval 90.6-100 %). Survival for the endpoint aseptic loosening at 2 years was 100 %. The HHS (Harris Hip Score) improved from 56 (14-86) preoperatively to 95 (60-100) 2 years after the operation. Cementing results were judged 47 % Grade A, 45 % Grade B and 7 % Grade C. Osteolysis was found in 2 stems without clinical symptoms or correlation to subsidence or cementing quality. The EBRA-FCA analysis showed an average subsidence of -0.30 mm (95 % CI -0.5 mm to -0.1 mm). 11 patients showed a subsidence of more than 1 mm. In this group one patient showed a subsidence of 1.5 mm and one of 3.1 mm without further radiological changes. CONCLUSIONS The twinSys stem showed excellent clinical and radiologic short term results at 2 years follow-up and seems to be a reliable implant.
منابع مشابه
Medium-term Outcome of Cemented Titanium Stems
Early failures have seen titanium fall from favour as a material for cemented femoral stems. Between 1989 and 1994, we performed a prospective review of a consecutive series of 122 cemented total hip replacements using the Ultima straight textured titanium stem, and report the fiveto ten-year clinical and radiological outcomes. There were no revisions for loosening of the femoral stem. Revision...
متن کاملStem subsidence of polished and rough double-taper stems
BACKGROUND AND PURPOSE Many clinical reports have indicated that polished hip stems show better clinical results than rough stems of the same geometry. It is still unknown, however, what the mechanical effects are of different surface finishes on the cement at the cement-bone interface. We compared mechanical effects in an in vitro cemented hip arthroplasty model. METHODS Two sizes of double-...
متن کاملCollarless polished tapered stem: clinical and radiological results at a minimum of ten years' follow-up.
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cem...
متن کاملFive-year prospective clinical and radiological results of a new cannulated cemented polished Tri-Taper femoral stem.
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different ...
متن کاملLong-term Outcome of Polished Stems in Total Hip Arthroplasty
PURPOSE The purpose of this study was to evaluate the clinical outcomes and radiologic findings of primary total hip arthroplasty (THA) using cemented polished femoral stems. MATERIALS AND METHODS We retrospectively reviewed 91 hips (91 patients) that had undergone primary THA with cemented polished femoral stems who were followed for at least 10 years. The mean age at surgery was 57 years (4...
متن کامل