The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices
نویسندگان
چکیده
PURPOSE This study aimed to assess the combined effect of dismantling before sterilization and aging on the accuracy (±10% of the target torque) of spring-style mechanical torque devices (S-S MTDs). METHODS Twenty new S-SMTDs from two different manufacturers (Nobel Biocare and Straumann: 10 of each type) were selected and divided into two groups, namely, case (group A) and control (group B). For sterilization, 100 cycles of autoclaving were performed in 100 sequences. In each sequence, 10 repetitions of peak torque values were registered for aging. To measure and assess the output of each device, a Tohnichi torque gauge was used (P<0.05). RESULTS Before steam sterilization, all of the tested devices stayed within 10% of their target values. After 100 cycles of steam sterilization and aging with or without dismantling of the devices, the Nobel Biocare devices stayed within 10% of their target torque. In the Straumann devices, despite the significant difference between the peak torque and target torque values, the absolute error values stayed within 10% of their target torque. CONCLUSION Within the limitations of this study, there was no significant difference between the mean and absolute value of error between Nobel Biocare and Straumann S-S MTDs.
منابع مشابه
The effect of steam sterilization on the accuracy of spring-style mechanical torque devices for dental implants
BACKGROUND Mechanical torque devices (MTDs) are one of the most commonly recommended devices used to deliver optimal torque to the screw of dental implants. Recently, high variability has been reported about the accuracy of spring-style mechanical torque devices (S-S MTDs). Joint stability and survival rate of fixed implant supported prosthesis depends on the accuracy of these devices. Currentl...
متن کاملEffect of Component Dismantling on Quality of Sterilization of Mechanical Torque Limiting Devices(Pilot Study)
Objectives Dental instruments may serve as a route of infection transmission. The necessity of dismantling of Friction-Style (FS) and SpringStyle (SS) mechanical torque limiting devices (MTLDs) prior to disinfection and sterilization is a matter of debate, aside fromthe fact that it is difcult and time-consuming. This study sought to assess the effect of dismantling of components on quality of ...
متن کاملبررسی تأثیر دفعات کارکرد بر دقت محدودکنندههای مکانیکی تورک از نوع فنری در ایمپلنتهای دندانی□
Objective: Recently, a wide variation has been reported in delivery of the target torque by spring-style mechanical torque limiting devices (MTLDs) but the effect of aging on their accuracy has not been independently evaluated. The purpose of this study was to determine the effect of aging on the accuracy of spring-style MTLDs. Methods: In this in-vitro experimental study, 15 new spring-style M...
متن کاملEvaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
OBJECTIVE Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization process...
متن کاملThe effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
BACKGROUND Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devic...
متن کامل