Evaluation of manual resuscitators used in ICUs in Brazil*

نویسندگان

  • Tatiana de Arruda Ortiz
  • Germano Forti
  • Márcia Souza Volpe
  • Marcelo do Amaral Beraldo
  • Marcelo Britto Passos Amato
  • Carlos Roberto Ribeiro Carvalho
  • Mauro Roberto Tucci
چکیده

OBJECTIVE To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. METHODS Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting valve; and tidal volume (VT) generated when the one-handed and two-handed techniques were used. The tests were performed and analyzed in accordance with the American Society for Testing and Materials (ASTM) F920-93 criteria. RESULTS Expiratory resistance was greater than 6 cmH2O . L-1 . s-1 in only one MR. The pressure-limiting valve, a feature of five of the MRs, opened at low pressures (< 17 cmH2O), and the maximal pressure was 32.0-55.9 cmH2O. Mean VT varied greatly among the MRs tested. The mean VT values generated with the one-handed technique were lower than the 600 mL recommended by the ASTM. In the situations studied, mean VT was generally lower from the Brazilian-made MRs that had a pressure-limiting valve. CONCLUSIONS The resistances imposed by the patient valve met the ASTM criteria in all but one of the MRs tested. The pressure-limiting valves of the Brazilian-made MRs usually opened at low pressures, providing lower VT values in the situations studied, especially when the one-handed technique was used, suggesting that both hands should be used and that the pressure-limiting valve should be closed whenever possible.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of 16 adult disposable manual resuscitators.

INTRODUCTION Disposable manual resuscitators are commonly used to ventilate patients during cardiopulmonary resuscitation, suctioning, and intrahospital transport, and their clinical performance is critical. METHODS We bench-tested 16 adult disposable manual resuscitators from 9 different manufacturers. We performed a series of tests and made observations using testing industry standards as a...

متن کامل

Comparison of the FiO2 delivered by seven models of the self-inflating bag-mask system.

BACKGROUND AND OBJECTIVES Since resuscitators with self-inflating bag-mask systems manufactured and/or commercialized in Brazil are widely available and used in health services, both out- and intra-hospitals, the objective of this study was to determine the O2 fractions delivered by seven resuscitators receiving different O2 flows. METHODS Seven resuscitators with self-inflating bag-mask syst...

متن کامل

Efficiency of manual resuscitators at birth.

The effectiveness of five neonatal/paediatric manual resuscitators was assessed in a group of babies born by caesarean section. Results showed that devices incorporating a large volume reservoir produced the greatest tidal volume, while those with smaller volume reservoirs could not be considered satisfactory for routine use during neonatal resuscitation.

متن کامل

Performance and Acceptability of Two Self-Inflating Bag-Mask Neonatal Resuscitator Designs.

BACKGROUND A self-inflating bag-mask device is specified by international policy guidelines as standard prototype of care for newborn resuscitation. Our hypothesis is that a new bag-mask design would be as effective and easy to use as a standard, self-inflating resuscitation bag-mask. METHODS We conducted a comparative evaluation of the performance and acceptability of the Laerdal 220-mL resu...

متن کامل

مقایسه بکارگیری دو روش تمرینات تنفسی غیر فعال دق دستی و ارتعاش مکانیکی بر عملکرد قلبی و ریوی بیماران تحت تهویه مکانیکی در بخش مراقبت‌های ویژه

Introduction & Objective: passive respiratory exercises are used to prevent from accumulation of secretions, lung collapse and the incidence of ventilator-associated pneumonia in ICUs. This  study is aimed to compare the effects of manual percussion and mechanical vibrations on cardiopulmonary functions of mechanically ventilated patients. Materials &Methods: This randomized crossover clinical...