A Comparison of Three Dimensional Ultrasound, Clips and CT for Measuring Interfractional Breast Lumpectomy Cavity Motion

نویسندگان

  • Ruth Heimann
  • Daphne Hard
  • Jessica Archambault
  • Suzanne Gross
  • Martin Lachaine
چکیده

Accurate definition of the lumpectomy cavity (LC) is essential for both partial breast and electron boost planning and delivery. The goal of this study is to evaluate CT, 3DUS and clips in the daily target localization of the LC. Twenty whole breast radiation patients, including 10 with clips, underwent two sets of treatment planning CTs (CT1, CT2) with co-registered 3DUS (US1, US2), separated by approximately 6 weeks. The cavities were independently outlined on each CT and 3DUS dataset. Of the 20 patients, 18 had visible cavities. The shifts in cavity position were calculated from CT, 3DUS and clips. The PTV margins required to be added to the CT1 cavity volume in order to fully encompass the CT2 cavity were calculated without and with shifts according to 3DUS or clips. The median cavity displacement ranged from 0.0-3.0 mm with a maximum of up to 17.9 mm. The cavity displacements were not significantly different when calculated based on 3DUS, CT, or clips (p > 0.05). The average PTV margin required to encompass the CT2 cavity was 6.3 mm without shifts, 4.3 mm with 3DUS shifts and 5.2 mm with clip-based shifts. The differences in the margin between no shift and 3DUS or clip-guided shift were found to be significant, while differences between 3DUS and clip-based shifts were not (p > 0.7). With no shift, the mean CT2 volume that lied outside the CT1 was 3.2 cc compared to 2.4 cc (p < 0.03) with 3DUS and 2.8 cc (p < 0.05) with clip-based shift. The volume difference between the shifting methods was not significant (p > 0.9). Without adjusting for the changes in the cavity shape and location, a portion of the volume will be undertreated unless the margins are increased. 3DUS may offer an easily implemented solution to localizing the LC without requiring additional ionizing radiation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ultrasound-guided boost irradiation of tumor cavity after lumpectomy in breast cancer

Background: &nbsp;After breast conserving surgery, most recurrences occur around the primary tumor site. This justifies the use of boost radiotherapy to the primary site of tumor. Surgical scar is not always a good surrogate for the location of the lumpectomy cavity. The aim of this study was to evaluate ultrasonic guidance for detection of the lumpectomy cavity after breast conserving surgery ...

متن کامل

Displacement of Surgical Clips during Postoperative Radiotherapy in Breast Cancer Patients Who Received Breast-Conserving Surgery

PURPOSE Surgical clips are used as a target for postoperative breast radiotherapy, and displacement of surgical clips would result in inaccurate delivery of radiation. We investigated the displacement range of surgical clips in the breast during postoperative radiotherapy following breast-conserving surgery. METHODS A total of 178 patients who received breast-conserving surgery and postoperat...

متن کامل

3D-conformal Accelerated Partial Breast Irradiation treatment planning: the value of surgical clips in the delineation of the lumpectomy cavity

BACKGROUND Accurate localisation of the lumpectomy cavity (LC) volume is one of the most critical points in 3D-conformal Partial breast irradiation (3D-APBI) treatment planning because the irradiated volume is restricted to a small breast volume. Here, we studied the role of the placement of surgical clips at the 4 cardinal points of the lumpectomy cavity in target delineation. METHODS Forty ...

متن کامل

Dosimetric Comparison of 3D Tangential Radiotherapy of Post-Lumpectomy Breast at Two Different Energies

Introduction: Radiation therapy following breast conserving surgery is one of the most common procedures performed in any radiation oncology department. A tangential parallel-opposed pair is almost always the technique of choice for this purpose. This technique is often performed based on 3D treatment planning. The aim of this study was to compare 3D treatment planning for two different energie...

متن کامل

Image guidance using 3D-ultrasound (3D-US) for daily positioning of lumpectomy cavity for boost irradiation

PURPOSE The goal of this study was to evaluate the use of 3D ultrasound (3DUS) breast IGRT for electron and photon lumpectomy site boost treatments. MATERIALS AND METHODS 20 patients with a prescribed photon or electron boost were enrolled in this study. 3DUS images were acquired both at time of simulation, to form a coregistered CT/3DUS dataset, and at the time of daily treatment delivery. I...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2016