A Readout IC Using Two-Step Fastest Signal Identification for Compact Data Acquisition of PET Systems
نویسندگان
چکیده
A readout integrated circuit (ROIC) using two-step fastest signal identification (FSI) is proposed to reduce the number of input channels of a data acquisition (DAQ) block with a high-channel reduction ratio. The two-step FSI enables the proposed ROIC to filter out useless input signals that arise from scattering and electrical noise without using complex and bulky circuits. In addition, an asynchronous fastest signal identifier and a self-trimmed comparator are proposed to identify the fastest signal without using a high-frequency clock and to reduce misidentification, respectively. The channel reduction ratio of the proposed ROIC is 16:1 and can be extended to 16 × N:1 using N ROICs. To verify the performance of the two-step FSI, the proposed ROIC was implemented into a gamma photon detector module using a Geiger-mode avalanche photodiode with a lutetium-yttrium oxyorthosilicate array. The measured minimum detectable time is 1 ns. The difference of the measured energy and timing resolution between with and without the two-step FSI are 0.8% and 0.2 ns, respectively, which are negligibly small. These measurement results show that the proposed ROIC using the two-step FSI reduces the number of input channels of the DAQ block without sacrificing the performance of the positron emission tomography (PET) systems.
منابع مشابه
Development of compact DOI-measurable PET detectors for simultaneous PET/MR Imaging
University of Texas MD Anderson Cancer Center, USA It is critically needed yet challenging to develop compact PET detectors with high sensitivity and uniform, high imaging resolution for improving the performance of simultaneous PET/MR imaging, particularly for an integrated/inserted small-bore system. Using the latest “edge-less” SiPM arrays for DOI measurement using the design of dual-ended-s...
متن کاملFPGA based data acquistion and digital pulse processing for PET and SPECT
The most important aspects of nuclear medicine imaging systems such as Positron Emission Tomography (PET) or Single Photon Emission Computed Tomography (SPECT) are the spatial resolution and the sensitivity (detector efficiency in combination with the geometric efficiency). Considerable efforts have been spent during the last two decades in improving the resolution and efficiency by developing ...
متن کاملیک روش جستجوی دو- مرحلهای برای همزمانی سریع سیستمهای باند فوق وسیع در محیطهای چندمسیره
Ultra-wideband (UWB) radio is a promising technology for low-cost, high-rate communications at short ranges. However, high bandwidth and low power of the UWB signals impose some difficult challenges in signal processing and implementation of the receiver. Acquisition of the signal timing is the first task to be performed in a coherent receiver. The high bandwidth of UWB signals demand for fine ...
متن کاملSignal Design at lsolated lntersecaions Using Expert Systems Technology
The procedural steps for developing an expert system for designing signals at isolated intersections are described and the most important development issues for each step are discussed. The steps include problem analysis and definition, preliminary prototype specification, knowledge acquisition strategy development, prototype development plan, knowledge extraction, knowledge representation, too...
متن کاملEvaluation of Simultaneous Dual-radioisotope SPECT Imaging Using 18F-fluorodeoxyglucose and 99mTc-tetrofosmin
Objective(s): Use of a positron emission tomography (PET)/single-photonemission computed tomography (SPECT) system facilitates the simultaneousacquisition of images with fluorine-18 fluorodeoxyglucose (18F-FDG) andtechnetium (99mTc)-tetrofosmin. However, 18F has a short half-life, and 511keV Compton-scattered photons are detected in the 99mTc energy window.Therefore, in this study, we aimed to ...
متن کامل