Quality Assurance Procedures based on Dosimetric, Gamma Analysis as a Fast Reliable Tool for Commissioning Brachytherapy Treatment Planning Systems
نویسندگان
چکیده
Background Fast and easily repeatable methods for commissioning procedures for brachytherapy (BT) treatment planning systems (TPS) are needed. Radiochromic film dosimetry with gamma analysis is widely used in external beam quality assurance (QA) procedures and planar film dosimetry is also increasingly used for verification of the dose distribution in BT applications. Using the gamma analysis method for comparing calculated and measured dose data could be used for commissioning procedures of the newly developed TG-186 and MBDCA calculation algorithms. The aim of this study was dosimetric verification of the calculation algorithm used in TPS when the CT/MRI ring applicator is used. Materials and methods Ring applicators with 26 and 30 mm diameters and a 60 mm intra-uterine tube with 60° angle were used for verification. Gafchromic® EBT films were used as dosimetric media. Dose grids, corresponding to each plane (dosimetric film location), were exported from the TPS as a raw data. Gafchromic® films were digitized after irradiation. gamma analysis of the data were performed using the OMNI Pro I'mRT® system, as recommended by the AAPM TG-119 rapport criterion for gamma analysis of 3%, 3 mm and a level of 95%. Results For the 26 mm and 30 mm rings, the average gamma ranged, respectively, from 0.1 to 0.44 and from 0.1 to 0.27. In both cases, 99% of the measured points corresponded with the calculated data. Conclusions This analysis showed excellent agreement between the dose distribution calculated with the TPS and the doses measured by Gafchromic films. This finding confirms the viability of using film dosimetry in BT.
منابع مشابه
Developing a Verification and Training Phantom for Gynecological Brachytherapy System
Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...
متن کاملA Specific Patient Quality Assurance (PSQA) procedure for a Co-60 source based High Dose Rate Brachytherapy
Introduction: In radiation therapy, accurate dose determination and precise dose delivery to the tumor are directly associated with better treatment outcomes in terms of higher tumor control and lower post radiation therapy complications. The current study aims the development and clinical application of the Patient Specific Quality Assurance (PSQA) procedures for nasopharyngea...
متن کاملQuality assurance procedures during commissioning of a treatment planning system as a tool to establish new standards before migration
PURPOSE Treatment planning system commissioning is one of the most important parts of the quality assurance system in a working brachytherapy department. Migration to a more sophisticated system is always a step forward for the planning team but careful verification of the workflow and obtained results is mandatory. The question is not only whether the quality and safety of the previous standar...
متن کامل2D linear array device as a quality assurance tool in brachytherapy applications
Background: External beam radiotherapy and brachytherapy plays a vital role in the management of cancer cervix. High dose rate brachytherapy is being presently used worldwide for the brachytherapy applications. At present, 2-Dimensional linear array detectors are the most common QA tool used for pretreatment patient specific quality assurance in external beam radiotherapy alon...
متن کاملA novel phantom design for brachytherapy quality assurance
Background: One major challenge in brachytherapy is to verify the accuracy of dose distributions calculated by the treatment planning system. In this project, a new phantom design has been introduced for quality assurance of dose distributions in gynocological (GYN) brachytherapy implants using EBT GafChromic film. Materials and Methods: This phantom has been designed and fabricated from 90 sla...
متن کامل