Achieving an adequate minute volume through a 2 mm transtracheal catheter in simulated upper airway obstruction using a modified industrial ejector.
نویسندگان
چکیده
BACKGROUND Needle cricothyrotomy and subsequent transtracheal jet ventilation (TTJV) is one of the last options to restore oxygenation while managing an airway emergency. However, in cases of complete upper airway obstruction, conventional TTJV is ineffective and dangerous. We transformed a small, industrial ejector into a simple, manual ventilator providing expiratory ventilation assistance (EVA). METHODS An ejector pump was modified to allow both insufflation of oxygen and jet-assisted expiration through an attached 75 mm long transtracheal catheter (TTC) with an inner diameter (ID) of 2 mm by alternately occluding and releasing the gas outlet of the ejector pump. In a lung simulator, the modified ejector pump was tested at different compliances and resistances. Inspiration and expiration times were measured and achievable minute volumes (MVs) were calculated to determine the effect of EVA. RESULTS The modified ejector pump shortened the expiration time and an MV up to 6.6 litre min(-1) could be achieved through a 2 mm ID TTC in a simulated obstructed airway. CONCLUSIONS The principle of ejector-based EVA seems promising and deserves further evaluation.
منابع مشابه
Ventilation through a small-bore catheter: optimizing expiratory ventilation assistance.
BACKGROUND Emergency ventilation through a small-bore transtracheal catheter can be lifesaving in a 'cannot intubate, cannot ventilate' situation. Ejectors, capable of creating suction by the Bernoulli principle, have been proposed to facilitate expiration through small-bore catheters. In this bench study, we compared a novel, purpose-built ventilation ejector (DE 5) with a previously proposed,...
متن کاملTranstracheal ventilation with a novel ejector-based device (Ventrain) in open, partly obstructed, or totally closed upper airways in pigs.
BACKGROUND Transtracheal access and subsequent jet ventilation are among the last options in a 'cannot intubate-cannot oxygenate' scenario. These interventions may lead to hypercapnia, barotrauma, and haemodynamic failure in the event of an obstructed upper airway. The aim of the present study was to evaluate the efficacy and the haemodynamic effects of the Ventrain, a manually operated ventila...
متن کاملJet or intensive care unit ventilator during simulated percutaneous transtracheal ventilation: a lung model study.
BACKGROUND Percutaneous transtracheal ventilation (PTV) via a jet ventilator (PTJV) is considered a rescue technique in difficult airway management. However, whether a conventional ventilator can generate adequate ventilation via PTV is not known. Our goal was to evaluate the tidal volume (V(T)) generated by a conventional ventilator during simulated PTV compared with PTJV in a lung model. ME...
متن کاملRescue Ventilation Through a Small-Bore Transtracheal Cannula in Severe Hypoxic Pigs Using Expiratory Ventilation Assistance
BACKGROUND Suction-generated expiratory ventilation assistance (EVA) has been proposed as a way to facilitate bidirectional ventilation through a small-bore transtracheal cannula (TC). In this study, we investigated the efficiency of ventilation with EVA for restoring oxygenation and ventilation in a pig model of acute hypoxia. METHODS Six pigs (61-76 kg) were anesthetized and ventilated (int...
متن کاملPercutaneous Transtracheal Jet Ventilation with Various Upper Airway Obstruction
A "cannot-ventilate, cannot-intubate" situation is critical. In difficult airway management, transtracheal jet ventilation (TTJV) has been recommended as an invasive procedure, but specialized equipment is required. However, the influence of upper airway resistance (UAR) during TTJV has not been clarified. The aim of this study was to compare TTJV using a manual jet ventilator (MJV) and the oxy...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- British journal of anaesthesia
دوره 104 3 شماره
صفحات -
تاریخ انتشار 2010