Anatomy‐based, patient‐specific VMAT QA using EPID or MLC log files
نویسندگان
چکیده
In this project, we investigated the use of an electronic portal imaging device (EPID), together with the treatment planning system (TPS) and MLC log files, to determine the delivered doses to the patient and evaluate the agreement between the treatment plan and the delivered dose distribution. The QA analysis results are presented for 15 VMAT patients using the EPID measurements, the ScandiDos Delta4 dosimeter, and the beam fluence calculated from the multileaf collimator (MLC) log file. EPID fluence images were acquired in continuous acquisition mode for each of the patients and they were processed through an in-house MATLAB program to create an opening density matrix (ODM), which was used as the input fluence for the dose calculation in the TPS (Pinnacle3). The EPID used in this study was the aSi1000 Varian on a Novalis TX linac equipped with high-definition MLC. The actual MLC positions and gantry angles were retrieved from the MLC log files and the data were used to calculate the delivered dose distributions in Pinnacle. The resulting dose distributions were then compared against the corresponding planned dose distributions using the 3D gamma index with 3 mm/3% passing criteria. The ScandiDos Delta4 phantom was also used to measure a 2D dose distribution for all the 15 patients and a 2D gamma was calculated for each patient using the Delta4 software. The average 3D gamma using the EPID images was 96.1% ± 2.2%. The average 3D gamma using the log files was 98.7% ± 0.5%. The average 2D gamma from the Delta4 was 98.1% ± 2.1%. Our results indicate that the use of the EPID, combined with MLC log files and a TPS, is a viable method for QA of VMAT plans.
منابع مشابه
Investigation of error detection capabilities of phantom, EPID and MLC log file based IMRT QA methods
A patient specific quality assurance (QA) should detect errors that originate anywhere in the treatment planning process. However, the increasing complexity of treatment plans has increased the need for improvements in the accuracy of the patient specific pretreatment verification process. This has led to the utilization of higher resolution QA methods such as the electronic portal imaging devi...
متن کاملMonitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries.
This work investigated the differences between multileaf collimator (MLC) positioning accuracy determined using either log files or electronic portal imaging devices (EPID) and then assessed the possibility of reducing patient specific quality control (QC) via phantom-less methodologies. In-house software was developed, and validated, to track MLC positional accuracy with the rotational and sta...
متن کاملCorrelation of phantom‐based and log file patient‐specific QA with complexity scores for VMAT
The motivation for this study was to reduce physics workload relating to patient- specific quality assurance (QA). VMAT plan delivery accuracy was determined from analysis of pre- and on-treatment trajectory log files and phantom-based ionization chamber array measurements. The correlation in this combination of measurements for patient-specific QA was investigated. The relationship between del...
متن کاملSU-E-T-170: A Viable Approach to Patient Specific QA for Spine VMAT SRS Using EPID-Based Dosimetry.
PURPOSE VMAT involves multi-parameter modulations, thus presenting multi-dimensional challenges. Here, we present a viable approach to VMAT patient specific QA using EPID-based dosimetry for spine VMAT SRS casesMethods: Our spine VMAT SRS plans consisted of two full arcs and were optimized with 6 MV photons for a Varian Trilogy LINAC. The QA plans were computed by resetting all control points t...
متن کاملAn EPID‐based system for gantry‐resolved MLC quality assurance for VMAT
Multileaf collimator (MLC) positions should be precisely and independently mea-sured as a function of gantry angle as part of a comprehensive quality assurance (QA) program for volumetric-modulated arc therapy (VMAT). It is also ideal that such a QA program has the ability to relate MLC positional accuracy to patient-specific dosimetry in order to determine the clinical significance of any dete...
متن کامل