Monte Carlo calculations of electron beam quality conversion factors for several ion chamber types.

نویسندگان

  • B R Muir
  • D W O Rogers
چکیده

PURPOSE To provide a comprehensive investigation of electron beam reference dosimetry using Monte Carlo simulations of the response of 10 plane-parallel and 18 cylindrical ion chamber types. Specific emphasis is placed on the determination of the optimal shift of the chambers' effective point of measurement (EPOM) and beam quality conversion factors. METHODS The EGSnrc system is used for calculations of the absorbed dose to gas in ion chamber models and the absorbed dose to water as a function of depth in a water phantom on which cobalt-60 and several electron beam source models are incident. The optimal EPOM shifts of the ion chambers are determined by comparing calculations of R50 converted from I50 (calculated using ion chamber simulations in phantom) to R50 calculated using simulations of the absorbed dose to water vs depth in water. Beam quality conversion factors are determined as the calculated ratio of the absorbed dose to water to the absorbed dose to air in the ion chamber at the reference depth in a cobalt-60 beam to that in electron beams. RESULTS For most plane-parallel chambers, the optimal EPOM shift is inside of the active cavity but different from the shift determined with water-equivalent scaling of the front window of the chamber. These optimal shifts for plane-parallel chambers also reduce the scatter of beam quality conversion factors, kQ, as a function of R50. The optimal shift of cylindrical chambers is found to be less than the 0.5 rcav recommended by current dosimetry protocols. In most cases, the values of the optimal shift are close to 0.3 rcav. Values of kecal are calculated and compared to those from the TG-51 protocol and differences are explained using accurate individual correction factors for a subset of ion chambers investigated. High-precision fits to beam quality conversion factors normalized to unity in a beam with R50 = 7.5 cm (kQ (')) are provided. These factors avoid the use of gradient correction factors as used in the TG-51 protocol although a chamber dependent optimal shift in the EPOM is required when using plane-parallel chambers while no shift is needed with cylindrical chambers. The sensitivity of these results to parameters used to model the ion chambers is discussed and the uncertainty related to the practical use of these results is evaluated. CONCLUSIONS These results will prove useful as electron beam reference dosimetry protocols are being updated. The analysis of this work indicates that cylindrical ion chambers may be appropriate for use in low-energy electron beams but measurements are required to characterize their use in these beams.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo calculations of beam quality correction factors

Current dosimetry protocols (AAPM, IAEA, DIN) recommend the use of parallel-plate ionization chambers for the measurement of absorbed dose-towater in clinical electron beams. For well-guarded plane-parallel chambers, it is assumed that the perturbation correction pQ is unity for all electron energies. In this study, we present detailed Monte Carlo simulations with the EGSnrc code for the widely...

متن کامل

Measurements of Photon Beam Flattening Filter Using an Anisotropic Analytical Algorithm and Electron Beam Employing Electron Monte Carlo

  Introduction: This study aimed to report the measurement of photon and electron beams to configure the Analytical Anisotropic Algorithm and Electron Monte Carlo used in clinical treatment. Material and Methods: All measurements were performed in a large water phantom using a 3-dimensional scanning system (PTW, Germany). ...

متن کامل

Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation

Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...

متن کامل

An Efficiency Studying of an Ion Chamber Simulation Using Vriance Reduction Techniques with EGSnrc

Background: Radiotherapy is an important technique of cancer treatment using ionizing radiation. The determination of total dose in reference conditions is an important contribution to uncertainty that could achieve 2%. The source of this uncertainty comes from cavity theory that relates the in-air cavity dose and the dose to water. These correction factors are determined from Monte Carlo calcu...

متن کامل

Radiation measurements in secondary elctron beam and comparison with calculations.

A high energy particle beamline is described which is capable of transporting a high intensity electron beam to potentially occupied areas. Ionization chamber measurements of electron beam intensity are described together with the ion chamber calibration and necessary corrections. The results of the measurements are compared with a Monte Carlo calculation of the electron intensity. (Submitted t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 41 11  شماره 

صفحات  -

تاریخ انتشار 2014