Chemiluminescence immunoassay of thyrotropin with acridinium-ester-labeled antibody evaluated and compared with two other immunoassays.

نویسندگان

  • M P Bounaud
  • J Y Bounaud
  • M H Bouin-Pineau
  • L Orget
  • F Begon
چکیده

A new chemiluminometric immunoassay of thyrotropin (TSH) involves antibody labeled with acridinium ester ("Magic Lite System," Ciba Corning Diagnostic Corp.). The assay is rapid, with two incubations totaling 2.5 h, requires two standards per run, and takes 10 s per sample for the quantification step. Analytical performance, within- and between-run reproducibilities, and linearity were excellent. The detection limit is 0.04 milli-int. unit/L. Results correlated well with those obtained by immunoradiometric assay (RIA-gnost hTSH, Hoechst-Behring) and immunofluorometric assay (hTSH Delfia, LKB): r = 0.975. TSH measurements in 32 euthyroid subjects ranged from 0.4 to 4.8 milli-int. units/L (mean 1.35 milli-int. units/L). TSH values for 51 hypothyroid and subclinically hypothyroid patients ranged from 2 to 65 milli-int. units/L. TSH values for 33 hyperthyroid patients (less than 0.14 milli-int. unit/L, less than 0.04 milli-int. unit/L in 16 of the 33) were clearly lower than for most untreated euthyroid subjects. For 169 other individuals whose thyroid function was being routinely assessed. TSH ranged from 0.4 to 4.8 milli-int. units/L, three had TSH less than 0.14 milli-int. unit/L, and four had TSH between 0.14 and 0.4 milli-int. unit/L. This system is as efficient and reliable for screening for thyroid function as the two comparison systems.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Synthesis and properties of differently charged chemiluminescent acridinium ester labels.

Chemiluminescent acridinium dimethylphenyl esters containing N-sulfopropyl groups in the acridinium ring are highly sensitive, hydrophilic labels that are used in automated immunoassays for clinical diagnostics. Light emission from these labels is triggered with alkaline peroxide in the presence of a cationic surfactant. At physiological pH, N-sulfopropyl acridinium esters exist as water adduct...

متن کامل

Preparation of an Acridinium Ester-Labeled Antibody and Its Application in GoldMag Nanoparticle-Based, Ultrasensitive Chemiluminescence Immunoassay for the Detection of Human Epididymis Protein 4

An ultrasensitive and rapid sandwich-type chemiluminescence immunoassay (CLIA) was developed for the clinical determination of human epididymis protein 4 (HE4) in human serum, using GoldMag nanoparticles as solid phase and acridinium ester (AE) as chemiluminescence system (GMP-CLIA). The process of AE labeling antibodies was systematically studied and evaluated. The effect of varies factors suc...

متن کامل

Chemiluminescence from alkoxy-substituted acridinium dimethylphenyl ester labels.

Chemiluminescent acridinium dimethylphenyl ester labels are used in automated immunoassays for clinical diagnostics. Light emission from these labels is triggered by alkaline peroxide in the presence of the cationic surfactant cetyltrimethylammonium chloride (CTAC). The surfactant plays a critical role in the chemiluminescence process of these labels by both accelerating their emission kinetics...

متن کامل

The Study of Chemiluminescence of Acridinium Ester in Presence of Rhodamin B as a Fluorescer

The reaction of hydrogen peroxide and  pheny -9-acridinecarboxylate in the presence of rhodamin B as a fluorophore leads to the emission of either bright blue light  or bright red light, depending on the structure of fluorescer and reaction conditions. The light emitter responsible for the Chemi Luminescence (CL) reaction is acridone and that the blue light produced with out the fluorescers...

متن کامل

Chemiluminescent labeled-antibody assay for thyroxin in serum, with magnetic separation of the solid-phase.

A chemiluminescent labeled-antibody immunoassay has been developed for measurement of total thyroxin (T4) in serum. Monoclonal antibodies to T4 labeled with a chemiluminescent acridinium ester are used. Serum samples are incubated with the labeled antibodies and a thyroxin-rabbit immunoglobulin conjugate, then reacted with magnetizable particles containing sheep anti-rabbit immunoglobulin. The ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 33 11  شماره 

صفحات  -

تاریخ انتشار 1987