Modified point diffraction interferometer for inspection and evaluation of ophthalmic components.

نویسندگان

  • Eva Acosta
  • Sara Chamadoira
  • Ralf Blendowske
چکیده

We demonstrate that a modified point diffraction interferometer can be used to measure the power distribution of different kinds of ophthalmic lenses such as spectacles, rigid and soft contact lenses, progressive lenses, etc. The relationship between the shape of the fringes and the power characteristics of the component being tested is simple and makes the design a very convenient and robust tool for inspection or quality control. Some simulations based on the Fresnel approximation are included.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Infrared point diffraction interferometer.

The point-diffraction interferometer (PDI) is a simple common-path interferometer capable of measuring optical-path differences (OPD) directly. This interferometer, first described by Smartt in 1972,1,2 has since become available commercially for use in the visible spectrum. Currently activity in laser-fusion and high-energy-laser systems has stimulated interest in extending PDI principles to t...

متن کامل

A Novel Method for Ultrasonic Evaluation of Horizontal Defects Using Time-of-Flight Diffraction

Time-of-flight diffraction method (ToFD) is an amplitude-independent sizing method which is based on the measurement of time-of-flight of defect tip diffracted waves. Although ToFD can measure through-wall length of defect accurately, this method is not capable of measuring horizontal defect size. In this paper, a new ToFD method for evaluating horizontal planar defects is presented. The finite...

متن کامل

Analysis of Frequency Leakage in Different Optical Paths of Nano-Metrology Systems Based on Frequency-Path Models

The drawing of frequency-path (F-P) models of optical beams is an approach for nonlinearity analysis in nano-metrology systems and sensors based on the laser interferometers. In this paper, the frequency-path models of four nano-metrology laser interferometry systems are designed, analyzed and simulated, including ...

متن کامل

Phase-shifting point diffraction interferometer.

We describe a novel interferometer design suitable for highly accurate measurement of wave-front aberrations over a wide range of wavelengths, from visible to x ray. The new design, based on the point diffraction interferometer, preserves the advantages of the conventional point diffraction interferometer but offers higher efficiency and improved accuracy through phase shifting. These qualities...

متن کامل

Fourier transform interferometer alignment method.

A rapid and convenient method has been developed to facilitate the alignment of the image-plane components of point-diffraction interferometers, including the phase-shifting point-diffraction interferometer. In real time, the Fourier transform of the detected image is used to calculate a pseudoimage of the electric field in the image plane of the test optic where thecritical alignment o f vario...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of the Optical Society of America. A, Optics, image science, and vision

دوره 23 3  شماره 

صفحات  -

تاریخ انتشار 2006