Finite Element Study of Matched Paired Posterior Disc Implant and Dynamic Stabilizer (360° Motion Preservation System)

نویسندگان

  • Vijay K. Goel
  • Ali Kiapour
  • Ahmed Faizan
  • Manoj Krishna
  • Tai Friesem
چکیده

BACKGROUND Anterior lumbar disc replacements are used to restore spinal alignment and kinematics of a degenerated segment. Compared to fusion of the segment, disc replacements may prevent adjacent segment degeneration. To resolve some of the deficiencies of anterior lumbar arthroplasty, such as the approach itself, difficulty of revision, and postoperative facet pain, 360° motion preservation systems based on posterior disc and posterior dynamic system (PDS) designs are being pursued. These systems are easier to revise and address all the pain generators in a motion segment, including the nerves, facets, and disc. However, biomechanics of the 360° posterior motion preservation system, including the contributions of the 2 subsystems (disc and PDS), are sparsely reported in the literature.nds. METHODS An experimentally validated 3-dimensional finite element model of the ligamentous L3-S1 segment was used to investigate the differences in biomechanical behavior of the lumbar spine. A single-level 360° posterior motion preservation system and its individual components in various orientations were simulated and compared with an intact model. Appropriate posterior surgical procedures were simulated. The PDS, a curved device with male and female components, was attached to the pedicle screws. The finite element models were subjected to 400 N of follower load plus 10Nm moment in extension and flexion. RESULTS The PDS restored flexion/extension motion to normal. The artificial disc led to increases in range of motion (ROM) compared with the intact model. ROM for the 360° system at the implanted and adjacent levels were similar to those of the respective intact levels. ROM was similar whether the discs were placed (a) both parallel to the midsagittal plane, (b) both angled 20° to the midsagittal plane, and (c) one at 20° and one parallel to the midsagittal plane. However, the stresses were slightly higher in the nonparallel disc configuration than in the parallel disc configuration, both in flexion and extension modes. CONCLUSIONS Posterior disc replacement with PDS restored the kinematics of the spine at all levels to near normal. In addition, placing the discs in a nonparallel configuration with respect to the midsagittal plane does not affect the functionality of the discs compared with parallel placement. Posterior disc replacement alone is not sufficient to restore the segment biomechanics to normal levels. CLINICAL RELEVANCE Finite element analysis results show that, unlike implants for fusion, PDS and posterior discs together (360° motion preservation system) are needed to preserve ROM. Such systems will prevent adjacent level degeneration and address pain from various spinal components, including facets.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Artificial Disc Position on Spine Biomechanics in the Cervical Spine: A Finite Element Study

INTRODUCTION Total disc replacement(TDR) has been suggested for surgical method to treat degenerative disc disease of the cervical spine because it decompresses nerves and preserves motion segments. The short-term outcomes of the TDR of cervical spine has been well reported to treat degenerative disc diseases as a motion preservation technique in spite of some clinical controversies such as imp...

متن کامل

Dynamic Load Carrying Capacity of Flexible Manipulators Using Finite Element Method and Pontryagin’s Minimum Principle

In this paper, finding Dynamic Load Carrying Capacity (DLCC) of flexible link manipulators in point to-point motion was formulated as an optimal control problem. The finite element method was employed for modelling and deriving the dynamic equations of the system. The study employed indirect solution of optimal control for system motion planning. Due to offline nature of the method, many diffic...

متن کامل

Anatomic Facet Replacement System (AFRS) Restoration of Lumbar Segment Mechanics to Intact: A Finite Element Study and In Vitro Cadaver Investigation

BACKGROUND Many decompression procedures involve complete or partial facetectomy. Spinal fusion usually stabilizes the motion segment after complete facetectomies. However, problems with fusion, such as adjacent-level degeneration, have increased interest in motion- preservation technologies. Facet arthroplasty may become an important posterior motion-preservation device, but its biomechanical ...

متن کامل

Biomechanical Evaluation of Pedicle Screw-Based Dynamic Stabilization Devices for the Lumbar Spine: A Systematic Review

STUDY DESIGN This study is a systematic review of published biomechanical studies involving pedicle screw-based posterior dynamic stabilization devices (PDS) with a special focus on kinematics and load transmission through the functional spine unit (FSU). METHODS A literature search was performed via the PubMed online database from 1990 to 2008 using the following key words: "biomechanics," "...

متن کامل

Biomechanical evaluation of a new pedicle screw-based posterior dynamic stabilization device (Awesome Rod System) - a finite element analysis

BACKGROUND Pedicle-screw-based posterior dynamic stabilization devices are designed to alleviate the rate of accelerated degeneration of the vertebral level adjacent to the level of spinal fusion. A new pedicle-screw-based posterior dynamic stabilization device- the Awesome Dynamic Rod System was designed with curve cuts on the rods to provide flexibility. The current study was conducted to eva...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2007