Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene
نویسندگان
چکیده
More than 60.000 hip arthroplasty are performed every year in Italy. Although Ultra-High-Molecular-Weight-Polyethylene remains the most used material as acetabular cup, wear of this material induces over time in vivo a foreign-body response and consequently osteolysis, pain, and the need of implant revision. Furthermore, oxidative wear of the polyethylene provoke several and severe failures. To solve these problems, highly cross-linked polyethylene and Vitamin-E-stabilized polyethylene were introduced in the last years. In in vitro experiments, various efforts have been made to compare the wear behavior of standard PE and vitamin-E infused liners. In this study we compared the in vitro wear behavior of two different configurations of cross-linked polyethylene (with and without the add of Vitamin E) vs. the standard polyethylene acetabular cups. The aim of the present study was to validate a micro X-ray computed tomography technique to assess the wear of different commercially available, polyethylene's acetabular cups after wear simulation; in particular, the gravimetric method was used to provide reference wear values. The agreement between the two methods is documented in this paper.
منابع مشابه
Quantification of Wear and Deformation in Different Configurations of Polyethylene Acetabular Cups Using Micro X-ray Computed Tomography
Wear is currently quantified as mass loss of the bearing materials measured using gravimetric methods. However, this method does not provide other information, such as volumetric loss or surface deviation. In this work, we validated a technique to quantify polyethylene wear in three different batches of ultrahigh-molecular-polyethylene acetabular cups used for hip implants using nondestructive ...
متن کاملWear Behaviours and Oxidation Effects on Different UHMWPE Acetabular Cups Using a Hip Joint Simulator
Given the long-term problem of polyethylene wear, medical interest in the new improved cross-linked polyethylene (XLPE), with or without the adding of vitamin E, has risen. The main aim of this study is to gain further insights into the mutual effects of radiation cross-linking and addition of vitamin E on the wear performance of ultra-high-molecular-weight polyethylene (UHMWPE). We tested four...
متن کاملWear of Crosslinked Uhmwpe in a Hip Joint Simulator
Crosslinking has been extensively introduced to reduce the wear of UHMWPE. In this study the wear rates and wear surfaces of UHMWPE with different levels of crosslinking were compared in a hip joint simulator. The UHMWPE was either non-irradiated, gamma irradiated in air (2.5MRad), or highly crosslinked (7.5 and 10MRad). The intentionally crosslinked materials showed a significantly lower volum...
متن کاملIn vivo response to cross-linked polyethylene and polycarbonate-urethane particles.
This study was undertaken to examine macrophage response to polycarbonate-urethane, a proposed alternative material to polyethylene in acetabular components of total hip arthroplasty. Polyethylene wear debris from total joint replacements has been linked to osteolysis and implant lifespan. It has been shown in vitro, that polyethylene particles cleaned of endotoxin generate less of an inflammat...
متن کاملKnee Simulator Wear of Cross-linked Uhmwpe with Normal and Mal-rotation Kinematics
Introduction: Wear of ultra-high molecular weight polyethylene (UHMWPE) is a major factor that affects longevity of the total joint replacement. In total hips (THR), cross-linking of polyethylene has been shown to be highly effective in reducing wear both clinically and experimentally [1-3]. Also, in TKR, Schmidig et al. [4] showed a 90% reduction of wear rate in 10 Mrad compared to 3 Mrad tibi...
متن کامل