Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments.

نویسندگان

  • Eugenio Pedullà
  • Nicola M Grande
  • Gianluca Plotino
  • Gianluca Gambarini
  • Ernesto Rapisarda
چکیده

INTRODUCTION The aim of this study was to evaluate the resistance to flexural fatigue of Reciproc R25 (VDW, Munich, Germany), WaveOne Primary (VDW, Munich, Germany), Mtwo (Sweden & Martina, Padova, Italy), and Twisted File (TF; SybronEndo, Orange, CA) instruments used in continuous rotation or in 2 different reciprocating motions. METHODS A total of 180 nickel-titanium files from 4 brands marketed, 2 for use in reciprocating motion (ie, Reciproc R25 and WaveOne Primary) and 2 for use in continuous rotation (ie, Mtwo and TF both taper .06/0.25 tip diameter), were tested. Forty-five instruments for each brand were divided into 3 groups (n = 15) on the basis of the motion tested: continuous rotation (group 1: 300 rpm) and reciprocal motion (group 2: "RECIPROC ALL" mode and group 3: "WAVEONE ALL" mode). The resistance to cyclic fatigue was determined by counting the number of cycles to failure in a 60° curve with a 5-mm radius. Data were analyzed through 2-way analyses of variance. RESULTS The cyclic fatigue resistance of the 2 reciprocating motion instruments (ie, "RECIPROC ALL" and "WAVEONE ALL") was significantly higher than the continuous rotation in each brand (P < .001). No significant difference was observed in cyclic fatigue between the 2 different reciprocal motions tested in each brand (P > .05). When considering the appropriate clinical motion for each brand, no significant difference in cyclic fatigue was found between TF, Mtwo, and Reciproc R25 instruments, whereas the cyclic fatigue resistance of WaveOne files was less than the other 3 brands (P < .05). CONCLUSIONS Reciprocal motion showed a significantly higher cyclic fatigue resistance in all brands compared with continuous rotation. No differences were found between the 2 reciprocating motions.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Influence of file motion on cyclic fatigue of new nickel titanium instruments.

AIM The aim of the present study was to compare the resistance to cyclic fatigue of Hyflex® (Coltene-Whaledent, Allstatten, Switzerland) size 40 taper.04 nickel titanium instruments used in continuous rotation versus the use with a reciprocating motion. The null hypothesis was that different file motions had no influence on the lifetime of instruments subjected to a cyclic fatigue test. METHO...

متن کامل

Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments?

AIM To compare cyclic fatigue resistance of two geometrically similar nickel-titanium instruments, used in conditions similar to clinical use in reciprocating and continuous rotary motion. METHODOLOGY Four groups of eighteen instruments each, Reciproc(®) files sizes ISO 25 and 40 (R25 and R40) and Mtwo(®) files sizes ISO 25 and 40 (M25 and M40), were tested in reciprocating and continuous rot...

متن کامل

TF TM Adaptive : a novel approach to nickel - titanium instrumentation

instruments have been traditionally used with a continuous rotary motion, but more recently, a new approach to the use of nickel-titanium instruments in a reciprocating movement has been introduced by Yared (2004). The clockwise (CW) and counter-clockwise (CCW) rotations used by Yared were four-tenths and two-tenths of a circle respectively, and the rotational speed utilized was 400 rpm. The co...

متن کامل

Resistance to flexural fatigue of Reciproc R25 files under continuous rotation and reciprocating movement.

INTRODUCTION The aim of the present work was to evaluate the resistance to flexural fatigue of Reciproc R25 nickel-titanium files, 25 mm, used in continuous rotation motion or reciprocation motion, in dynamic assays device. METHODS Thirty-six Reciproc R25 files were divided into 2 groups (n = 18) according to kinematics applied, continuous rotary (group CR) and reciprocation motion (group RM)...

متن کامل

Comparison of cyclic fatigue and torsional resistance in reciprocating single-file systems and continuous rotary instrumentation systems.

As compared with continuous rotary systems, reciprocating motion is believed to increase the fatigue resistance of NiTi instruments. We compared the cyclic fatigue and torsional resistance of reciprocating single-file systems and continuous rotary instrumentation systems in simulated root canals. Eighty instruments from the ProTaper Universal, WaveOne, MTwo, and Reciproc systems (n = 20) were s...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of endodontics

دوره 39 2  شماره 

صفحات  -

تاریخ انتشار 2013