Inter-instrumental agreement between goldman applanation tonometer (GAT) and non-contact tonometer (NCT) in relation to central corneal thickness (CCT) - an entity revisited
نویسندگان
چکیده
Background: To compare the level of intra ocular pressure measured by two different tonometers in relation to central corneal thickness (CCT). Material and Methods: Intraocular pressure (IOP) was measured in 118 consecutive patients first by Goldman applanation Tonometer (GAT) followed by Non-contact tonometer (NCT). Central corneal thickness was measured in each case by non-contact method. The values of IOP thus obtained by both the methods were compared in relation to the CCT values. Results: Out of total 118 patients, maximum subjects belonged to 31-40 year age-group (n=44). The mean age was 46.5 years. 66 male and 52 female participated in this study. The mean CCT was 517.68μm with standard deviation 26.77. 95 IOP readings were normal with NCT and 14 readings were graded as ocular hypertension (IOP measuring more than 21mm Hg) giving a specificity rate of 95.95% and sensitivity rate of 73.68%. Conclusion: In patients with normal range of CCT, GAT remains the method of choice.
منابع مشابه
Clinical comparison of Pulsair non-contact tonometer and Goldmann applanation tonometer in Indian population.
INTRODUCTION Goldmann applanation tonometer (GAT) is the gold standard for Intraocular Pressure (IOP) measurement but has disadvantage of being contact device and problems with portability. The aim of the study was to compare the Keeler's Pulsair noncontact tonometer (NCT) with GAT in Indian Population. MATERIALS AND METHODS Eighty-one subjects were screened from a Glaucoma clinic of a tertia...
متن کاملThe influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.
AIMS To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT). METHODS CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending...
متن کاملComparison of current tonometry techniques in measurement of intraocular pressure
PURPOSE To compare four tonometry techniques: Goldmann applanation tonometer (GAT), Dynamic contour tonometer (DCT), Non-contact tonometer (NCT), and Ocular Response Analyzer (ORA) in the measurement of intraocular pressure (IOP) and the impact of some corneal biomechanical factors on their performance. METHODS In this cross-sectional study, volunteers with normal ophthalmic examination and n...
متن کاملComparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects
BACKGROUND To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects. METHODS In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-530P, the Topcon CT-1P, and the GAT at the same time. We evaluated the IOP-mea...
متن کاملبررسی ارتباط بین ضخامت مرکز قرنیه و میزان فشار داخل چشمی اندازه گیری شده با تونومتر گلدمن
To determine the effect of central corneal thickness (CCT) on measured intraocular pressure (IOP) and evaluating the relationship between them in each of the groups of patients including: Normal subjects, ocular hypertension patients and patients with primary open angle glucoma (POAG). Correlation between intraocular pressure (Measured by Goldman applanation tonometer) and central corneal thick...
متن کامل