The influence of defect levels on the dose rate dependence of synthetic diamond detectors of various types on exposures to high-energy radiotherapy beams
نویسندگان
چکیده
The linear response of a radiation dosimeter with absorbed dose rate is a principal requirement in radiotherapy. Fowler's model for electrical conductivity, s of a solid-state detector and absorbed dose rate, Dr is of the form σ ∝ Δ Dr where Δ is the linearity index that can take on a range of values around unity. Utilising synthetic diamond detectors of various types as sensors, this study investigates the influence of defect levels on the Δ values of the sensors and the dependence ofΔ on bias voltage, beam energy and type in the dosimetry of highenergy photon and electron therapy beams. One main objective of the study was to establish whether for a given diamond detector, Δ could be determined only once for any given beam energy and then used for other beam energies of clinical interest. In order to attain the ICRU overall 75% uncertainty of absorbed dose delivery in radiotherapy, 72% accuracy was considered. The study was conducted on one HPHT and eight CVD synthesised diamonds of optical grade (OG) and detector grade (DG) qualities using 6 and 15MV photon, and 7 and 12 MeV electron energies. Values of Δ ranging from 0.79–1.03 to 0.85–0.96 were obtained for the electron and photon beams, respectively for all the diamond sensors at 1 kV/cm. The Δ values were found to change with various defect levels present within the crystals as characterised by Raman spectroscopy, ESR, FTIR spectroscopy and TL emission, and it was observed that the Δ values of crystals with high defect levels varied strongly with bias voltage. Whereas the Δ values of the HPHT diamond were found not vary with the electron and photon energies, only those of three CVD samples of a given class showed a variation within 2% between the two energies of each beam type. However, for all the crystals testedΔ showed amaximumvariation of 3.4% between the photon energies unlike the electron energies where a very strong variation (45%) was observed for three OG CVD crystals. The results of this study have suggested that differences in crystal quality due to the presence of defects could cause Δ to vary with bias voltage, beam energy and type. & 2014 Elsevier Ltd. All rights reserved.
منابع مشابه
Designing Optimal Bias Voltage for Radiotherapy Diamond Dosimeter
Introduction: Recent developments of radiotherapy techniques, require high accuracy detectors to determine the delivered dose in a small area. Chemical vapor deposition (CVD) or naturally growth diamond detectors which are commercially available are good candidates for this purpose. In these detectors two electrodes with high different electrical potential are deposited on both...
متن کاملEvaluating the effect of energy on calibration of thermo-luminescent dosimeters 7-LiF:Mg,Cu,P (GR-207A)
Background: 7-LiF: Mg,Cu,P (GR-207A) is one kind of thermo-luminescent dosimeters (TLD) which can be used in diagnostic radiology and radiotherapy applications. Because of having suitable properties such as good sensitivity, small size and being tissue equivalent, thermo-luminescent dosimetry has been one of the most important techniques in medicine over many years. Choosing a correct energy fo...
متن کاملAssessment of basic physical and dosimetric parameters of synthetic single-crystal diamond detector and its use in Leksell Gamma Knife and CyberKnife small radiosurgical fields
Background: To determine the basic physical and dosimetric properties of a new synthetic single-crystal diamond detector and its application for relative small field dosimetry. Materials and Methods: The pre-irradiation dose required to stabilize detector response, dose rate dependence, photon and electron energy dependence, temperature dependence and angular dependence of MicroDiamond detector...
متن کاملThesis Title : “ Evaluation of the Radiation Detection Properties of Synthetic
Aimed at improving the accuracy of dose determination in radiation medicine, this thesis explores the radiation detection properties of synthetic diamond crystals of various types and investigates the possibility of developing a single probe with synthetic diamond as sensor for the dosimetry of various beam types under large as well as small-field conditions. The study was conducted on two HPHT...
متن کاملVerification of dose rate and energy dependence of MAGICA polymer gel dosimeter with electron beams
Background: The purpose of this study was to evaluate the dependency of MAGICA polymer gel dosimeter response (R2) on different electron energies as well as on different mean dose rate for a standard clinically used linear accelerator. Materials and Methods: The sensitivity of the dosimeter was represented by the slope of calibration curve in the linear region measured for each modalit...
متن کامل