Perclose Closure Device breakage and embolization during deployment followed by retrieval with snare.

نویسندگان

  • Kevin Kang
  • Robert Ferraro
  • Richard Petrella
  • Gurjaipal Kang
چکیده

Perclose Closure Device (Abbott-Vascular) was attempted for femoral artery (FA) access site closure by an experienced operator. The device was felt to snap as it was pushed through the previously scarred skin and subcutaneous tissue. Perclose had broken into two pieces with the proximal portion outside the body and the distal portion embolizing to lie in the FA to distal aorta. The left FA was accessed through 8F sheath and 4-8mm Basket snare passed through a 6F multipurpose guiding catheter. The tip of the broken Perclose was caught with the snare and pulled out. The Perclose has a distal flexible sheath that is overmolded directly onto the proximal rigid guide. This junction of sheath and guide may be a weak point. Accidental pressure applied at this weak point in any Perclose device during deployment can cause breakage. Awareness of this rare complication is important for safety.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Urgent Surgical Intervention for Embolized Cardiac Occluder Devices: A Case Series

Introduction: In this study, we sought to illustrate our experience in urgent surgical management for embolized cardiac septal occlude devices resulting from trans-catheter closure of atrial septal defect and ventricular septal defect. Mathrials and Methods: We retrospectively reviewed four patients aged 2–10 years who underwent urgent surgery due to cardiac septal occluder embolization between...

متن کامل

Retrieval of Embolized Amplatzer Patent Foramen Ovale Occlusion Device: Issues Related to Late Recognition

Embolization of a percutaneous patent foramen ovale (PFO) closure device is a rare but serious complication. While early, periprocedural device embolization can normally be managed with snare and percutaneous retrieval, late embolization requires a different management strategy due to inability of the device to deform to allow passage into a large caliber sheath. We present a case of asymptomat...

متن کامل

Manual Aspiration Technique to Retrieve a Prematurely Detached Coil during Cerebral Aneurysm Embolization

Coil herniation, premature deployment, and fracture are procedure associated complications of endovascular aneurysm embolization that optimally necessitate coil retrieval when feasible. Several published techniques describe different strategies for managing coil complications including various snare retrieval devices, alligator retrieval devices, stent fixation, and open surgical resection of c...

متن کامل

Percutaneous transcatheter PDA device closure in infancy.

OBJECTIVE To evaluate the results and complications associated with transcatheter closure of patent ductus arteriosus (PDA) in infants. STUDY DESIGN Quasi-experimental study. PLACE AND DURATION OF STUDY Paediatric Cardiology Department of Armed Forces Institute of Cardiology / National Institute of Heart Diseases (AFIC/NIHD), Rawalpindi, from December 2010 to June 2012. METHODOLOGY Infant...

متن کامل

From the Southern Association for Vascular Surgery Can the Perclose suture-mediated closure system be used safely in patients undergoing diagnostic and therapeutic angiography to treat chronic lower extremity ischemia?

Purpose: Mechanical closure devices for arterial hemostasis after angiography, such as the Perclose suture-mediated closure system, are designed to decrease time to ambulation and improve patient comfort. Although these devices are safe and efficacious, to date there has been little reported about use of the Perclose device in a cohort consisting exclusively of patients with lower extremity per...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Indian heart journal

دوره 67 Suppl 3  شماره 

صفحات  -

تاریخ انتشار 2015