Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia.
نویسندگان
چکیده
PURPOSE To compare corneal pachymetry assessment using 4 measurement methods in eyes after laser in situ keratomileusis (LASIK) for myopia. SETTING Department of Ophthalmology & Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, Hong Kong SAR. METHODS Fifty-two consecutive patients (103 eyes) who had LASIK for the correction of myopia had Orbscan II (Bausch & Lomb), Visante (Carl Zeiss Meditec), Pentacam (Oculus, Inc.), and ultrasound (US) pachymetry (Sonomed, 200P) 6 months after surgery. Data were analyzed using the paired sample t test, Bland-Altman plots, and linear regression. RESULTS The mean postoperative pachymetry measured by US, Orbscan (0.89 acoustic factor), Pentacam, and Visante pachymetry were 438.2 microm+/-41.18 (SD), 435.17+/-49.63 microm, 430.66+/-40.23 microm, and 426.56+/-41.6 microm, respectively. Compared with the US measurement, Pentacam and Visante measurements significantly underestimated corneal thickness by a mean of 7.54+/-15.06 microm (P<.01) and 11.64+/-12.87 microm (P<.01), respectively. There was no statistically significant difference between US and Orbscan measurements. CONCLUSION Pentacam and Visante measurements of corneal thickness 6 months after LASIK were significantly less than those obtained using Orbscan and US pachymetry, although all 4 measurement methods showed a high correlation with each other.
منابع مشابه
Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy
PURPOSE To determine inter-device agreement for central corneal thickness (CCT) measurement among ultrasound pachymetry, rotating Scheimpflug imaging (Pentacam, Oculus, Wetzlar, Germany), and scanning slit corneal topography (Orbscan II, Bausch & Lomb, Rochester, NY, USA) in highly myopic eyes before and after photorefractive keratectomy (PRK). METHODS This prospective comparative study inclu...
متن کاملComparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.
AIMS To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). METHODS Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. RESULTS Three...
متن کاملSCIENTIFIC REPORT Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis
Aims: To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). Methods: Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. Results: Three ...
متن کاملComparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia.
PURPOSE To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post-laser in situ keratomileusis (LASIK) eyes for myopia. METHODS In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-LASIK patients using both the Pentacam and UP at the Eye Consultants Center...
متن کاملPosterior Corneal Curvature Assessment after Epi-LASIK for Myopia: Comparison of Orbscan II and Pentacam Imaging
PURPOSE To compare the changes in posterior corneal curvature using scanning slit topography (Orbscan II) and Scheimpflug imaging (Pentacam) before and after Epi-laser in situ keratomileusis (LASIK) for myopia. METHODS In a prospective observational case-series study, 20 myopic patients having undergone Epi-LASIK were examined serially with two different devices, Orbscan II and Pentacam, preo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of cataract and refractive surgery
دوره 33 7 شماره
صفحات -
تاریخ انتشار 2007