Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.
نویسندگان
چکیده
AIM To compare central corneal thickness (CCT) and anterior chamber depth (ACD) measurements using rotating Scheimpflug imaging and partial coherence interferometry. METHODS As part of the first phase of Shahroud Eye Cohort Study with 5 190 subjects of 40 to 64 years of age, CCT and ACD were measured using Scheimpflug imaging with the Pentacam (Oculus, Inc., Lynnwood, WA, USA) and partial coherence interferometry with the Allegro BioGraph (Wavelight, Erlangen, Germany). RESULTS After applying exclusion criteria, we had data of 4 387 subjects with a mean age of 50.7±6.2 years. Mean CCT with Pentacam and BioGraph were 528.6±33.2µm and 525.6±32µm respectively; the difference was statistically significant (P<0.001), but the correlation was high (R=0.920). Mean ACD measurements using Pentacam and BioGraph were 2.68±0.35mm and 2.62±0.33mm respectively; the inter-device difference was significant (P<0.001) with high correlation (R=0.944). The 95% limits of agreements between devices were -22.65µm to 28.61µm and -0.16mm to 0.29mm for CCT and ACD measurements, respectively. CONCLUSION For both CCT and ACD, the BioGraph gave significantly lower values than the Pentacam (P<0.05). Despite the high inter-device correlation, the 95% limits of agreements were wide, and this may limit their interchangeability in measuring the CCT and ACD.
منابع مشابه
Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts
Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 patients were included in this prospective study. The anterior chamber depth (ACD), keratometry (K), ...
متن کاملمقایسهی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبهی لنز داخل چشمی افراد با قرنیهی طبیعی
Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...
متن کاملComparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry.
PURPOSE To compare central corneal thickness (CCT), anterior chamber depth (ACD), and keratometry (K) readings measured using optical low-coherence reflectometry (OLCR) biometry and high-resolution rotating Scheimpflug photography. SETTING Eye Hospital of Wenzhou Medical College, Wenzhou, China. DESIGN Comparative case series. METHODS The CCT, ACD endothelium to lens, ACD epithelium to le...
متن کاملA comparison of Scheimpflug imaging simulated and Holladay equivalent keratometry values with partial coherence interferometry keratometry measurements in phakic eyes.
OBJECTIVE To evaluate the interchangeability of a Scheimpflug camera (Pentacam, Oculus) and a partial coherence interferometry keratometer (IOLMaster, Zeiss) for measures of keratometry and anterior chamber depth (ACD). A particular focus was to determine which Pentacam keratometry value best correlates with the IOLMaster and if these results can be used interchangeably in clinical practice. ...
متن کاملComparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
PURPOSE The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- International journal of ophthalmology
دوره 6 4 شماره
صفحات -
تاریخ انتشار 2013