CPAP titration by an auto-CPAP device based on snoring detection: a clinical trial and economic considerations.

نویسندگان

  • M Berkani
  • F Lofaso
  • C Chouaid
  • M Pia d'Ortho
  • D Theret
  • V Grillier-Lanoir
  • A Harf
  • B Housset
چکیده

This study aimed to assess the ability of an auto-nasal continuous positive airway pressure (nCPAP) device (REM + auto; NPBFD, Nancy, France) to predict the optimal constant nCPAP level. The apnoea/hypopnoea detection facility of the auto-nCPAP device was deliberately disabled and nasal mask pressure vibration detection was the only mode of pressure setting. The auto-nCPAP device was tested on 10 previously untreated patients with obstructive sleep apnoea during a single night, with ambulatory polysomnography performed in a conventional hospital room; the efficacy of the fixed pressure determined by the auto-nCPAP device was assessed by an ambulatory full polysomnography 2 weeks after the initiation of treatment at home. The fixed nCPAP pressure was effective (apnoea/hypopnoea and arousal indices <10 events x h(-1)) in all but two of the 10 patients studied. When the fixed nCPAP pressure was increased by 2 cmH2O in these two patients, sleep and respiration were normalized. Since only 12 ambulatory polysomnographic recordings were used to determine the effective nasal continuous positive airway pressure level, and as the device restored normal breathing and sleep in all 10 patients, it was concluded that this method of nasal continuous positive airway pressure titration may improve cost-effectiveness and reduce waiting lists in sleep laboratories.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Auto-CPAP: an effective and low-cost procedure in the management of OSAS?

Nasal continuous positive airway pressure (CPAP) is generally considered the first-line therapy in obstructive sleep apnoea (OSA) patients because it is effective at both correcting respiratory events and related arousals, and also improving morbidity [1]. Obstructive apnoeas and hypopnoeas can virtually always be eliminated by some level of CPAP <20 cmH2O [1, 2]. Determination of the prescript...

متن کامل

The use of auto-titrating continuous positive airway pressure for treatment of adult obstructive sleep apnea. An American Academy of Sleep Medicine review.

This paper reviews the efficacy of auto-titrating continuous positive airway pressure (APAP) for treatment of obstructive sleep apnea. It is based on a review of 30 articles published in peer review journals conducted by a task force appointed by the American Academy of Sleep Medicine to develop practice parameters for use of APAP devices for treatment of obstructive sleep apnea (OSA). The data...

متن کامل

Efficacy of automated continuous positive airway pressure in children with sleep-related breathing disorders in an attended setting.

INTRODUCTION The purpose of this study was to evaluate the safety and efficacy of automated continuous positive airway pressure (Auto-CPAP) in children. Sleep-related breathing disorders (SRBDs) include the clinical spectrum of symptomatic chronic snoring, upper airway resistance syndrome, and obstructive sleep apnea. This spectrum occurs in adults and children. Less data are available for chil...

متن کامل

Unattended home diagnosis and treatment of obstructive sleep apnea without polysomnography.

OBJECTIVE To test the effectiveness of unattended home monitoring along with automatic titrating continuous positive airway pressure (auto-CPAP) as an acceptable method for diagnosing and prescribing proper CPAP pressure for treatment of patients presenting with classic symptoms of obstructive sleep apnea (OSA). DESIGN Nonrandomized, prospective case study of 63 patients with a presumptive di...

متن کامل

Relative occurrence of flow limitation and snoring during continuous positive airway pressure titration.

OBJECTIVES To examine the relative temporal appearance of flow limitation and snoring during continuous positive airway pressure (CPAP) titration, compare their sensitivity as indicators of airway obstruction, and assess their relative utility as feedback variables for automatic titration of CPAP. DESIGN Retrospective review of data. SETTING University teaching hospital. PATIENTS Fifty-th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 12 4  شماره 

صفحات  -

تاریخ انتشار 1998