A novel implementation of mARC treatment for non-dedicated planning systems using converted IMRT plans
نویسندگان
چکیده
BACKGROUND The modulated arc (mARC) technique has recently been introduced by Siemens as an analogue to VMAT treatment. However, up to now only one certified treatment planning system supports mARC planning. We therefore present a conversion algorithm capable of converting IMRT plans created by any treatment planning system into mARC plans, with the hope of expanding the availability of mARC to a larger range of clinical users and researchers. As additional advantages, our implementation offers improved functionality for planning hybrid arcs and provides an equivalent step-and-shoot plan for each mARC plan, which can be used as a back-up concept in institutions where only one linac is equipped with mARC. METHODS We present a feasibility study to outline a practical implementation of mARC plan conversion using Philips Pinnacle and Prowess Panther. We present examples for three different kinds of prostate and head-and-neck plans, for 6 MV and flattening-filter-free (FFF) 7 MV photon energies, which are dosimetrically verified. RESULTS It is generally more difficult to create good quality IMRT plans in Pinnacle using a large number of beams and few segments. We present different ways of optimization as examples. By careful choosing the beam and segment arrangement and inversion objectives, we achieve plan qualities similar to our usual IMRT plans. The conversion of the plans to mARC format yields functional plans, which can be irradiated without incidences. Absolute dosimetric verification of both the step-and-shoot and mARC plans by point dose measurements showed deviations below 5% local dose, mARC plans deviated from step-and-shoot plans by no more than 1%. The agreement between GafChromic film measurements of planar dose before and after mARC conversion is excellent. The comparison of the 3D dose distribution measured by PTW Octavius 729 2D-Array with the step-and-shoot plans and with the TPS is well above the pass criteria of 90% of the points falling within 5% local dose and 3 mm distance to agreement. For all plans, the treatment time was noticeably reduced by conversion to mARC. CONCLUSIONS We present the feasibility test for converting IMRT step-and-shoot plans from the RTP-output of any treatment planning system (Philips Pinnacle and Prowess Panther, in our case) into mARC plans. The feasibility and dosimetric equivalence is demonstrated for the examples of a prostate and a head-and-neck patient.
منابع مشابه
mARC vs. IMRT radiotherapy of the prostate with flat and flattening-filter-free beam energies
BACKGROUND There as yet exists no systematic planning study investigating the novel mARC rotational radiotherapy technique, which is conceptually different from VMAT. We therefore present a planning study for prostate cancer, comparing mARC with IMRT treatment at the same linear accelerator equipped with flat and flattening-filter-free (FFF) photon energies. METHODS We retrospectively re-cont...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملPlanning for mARC treatments with the Eclipse treatment planning system
While modulated arc (mARC) capabilities have been available on Siemens linear accelerators for almost two years now, there was, until recently, only one treatment planning system capable of planning these treatments. The Eclipse treatment planning system now offers a module that can plan for mARC treatments. The purpose of this work was to test the module to determine whether it is capable of c...
متن کاملRadiobiological Model-Based Comparison of Three-Dimensional Conformal and Intensity-Modulated Radiation Therapy Plans for Nasopharyngeal Carcinoma
Introduction: Radiobiological modeling of radiotherapy plans are used for treatment plan comparisons. The current study aimed to compare the three-dimensional conformal radiation therapy (3DCRT) and intensity-modulated radiation therapy (IMRT) plans for nasopharyngeal cancer using radiobiological modeling. Materials and Methods: This study was conducted on 10 patients with nasopharyngeal carci...
متن کاملEvaluation of the effects of dental filling material artifacts on IMRT treatment planning in patient with nasopharyngeal cancer
Background: Presence of artifacts, caused by dental filling high-Z materials (DFM), on intensity-modulated radiation therapy (IMRT) treatment plan CT images may lead to uncertainty in head and neck calculated dose distributions. Hence, the purpose of this study was to investigate the effects of DFM on the IMRT calculated dose distribution and consequent radiobiological derived outcomes for naso...