A comparison of screw insertion torque and pullout strength.
نویسندگان
چکیده
PURPOSE Pullout strength of screws is a parameter used to evaluate plate screw fixation strength. However, screw fixation strength may be more closely related to its ability to generate sufficient insertion because stable nonlocked plate-screw fracture fixation requires sufficient compression between plate and bone such that no motion occurs between the plate and bone under physiological loads. Compression is generated by tightening of screws. In osteoporotic cancellous bone, sufficient screw insertion torque may not be generated before screw stripping. The effect of screw thread pitch on generation of maximum insertion torque (MIT) and pullout strength (POS) was investigated in an osteoporotic cancellous bone model and the relationship between MIT and POS was analyzed. METHODS Stainless steel screws with constant major (5.0 mm) and minor (2.7 mm) diameters but with varying thread pitches (1, 1.2, 1.5, 1.6, and 1.75 mm) were tested for MIT and POS in a validated osteoporotic surrogate for cancellous bone (density of 160 kg/m(3) [10 lbs/ft(3)]). MIT was measured with a torque-measuring hex driver for screws inserted through a one-third tubular plate. POS was measured after insertion of screws to a depth of 20 mm based on the Standard Specification and Test Methods for Metallic Medical Bone Screws (ASTM F 543-07). Five screws were tested for each failure mode and screw design. The relationship between MIT and compressive force between the plate and bone surrogate was evaluated using pressure-sensitive film. RESULTS There was a significant difference in mean MIT based on screw pitch (P < 0.0001), whereas POS did not show statistically significant differences among the different screw pitches (P = 0.052). Small screw pitches (1.0 mm and 1.2 mm) had lower MIT and were distinguished from large pitches (1.5 mm, 1.6 mm, and the 1.75 mm) with higher MIT. For POS, only the 1-mm and 1.6-mm pitch screws were found to be different from each other. Linear regression analysis of MIT revealed a moderate correlation to the screw pitch (R(2) = 0.67, P < 0.0001), whereas the analysis of POS suggested no correlation to the screw pitch (R(2) = 0.28, P = 0.006). Pearson correlation analysis indicated no correlation between MIT and POS (P = 0.069, r = -0.37). A linear relationship of increased compression between the plate and bone surrogate was found for increasing screw torque (R(2) = 0.97). CONCLUSIONS These results indicate that the ability of different screw designs to generate high screw insertion torque in a model of osteoporotic cancellous bone is unrelated to their pullout strength. Therefore, extrapolation of results for POS to identify optimal screw design for osteoporotic bone may not be valid. Screw designs that optimize MIT should be sought for fixation in osteoporotic bone.
منابع مشابه
Biomechanical Evaluation of the Influence of Cervical Screws Tapping and Design
OBJECTIVE To assess if the screw design (self-drilling/self-tapping) and the pilot hole tapping could affect the insertion torque and screw pullout strength of the screw used in anterior fixation of the cervical spine. METHODS Forty self-tapping screws and 20 self-drilling screws were inserted into 10 models of artificial bone and 10 cervical vertebrae of sheep. The studied parameters were th...
متن کاملInsertional torque and pullout strength of pedicle screws with or without repositioning: a porcine study.
PURPOSE. To evaluate the insertion torque and pullout strength of pedicle screws with or without repositioning. METHODS. 20 fresh porcine lumbar vertebrae of similar size were used. The entry point was at the site just lateral and distal to the superior facet joint of the vertebra, and to a depth of 35 mm. A 6.2-mm-diameter, 35-mm-long pedicle screw was inserted parallel to the superior end pla...
متن کاملEffect of Pilot Hole Tapping on Pullout Strength and Insertion Torque of Dual Core Pedicle Screws.
OBJECTIVE To evaluate the influence of pilot hole tapping on pullout resistance and insertion torque of pedicle screws with a conical core. METHODS Mechanical tests using a universal testing machine were performed on pedicle screws with a conical core that were inserted into pedicles in the fifth lumbar vertebra of calves. The insertion torque was measured using a torque meter with a capacity...
متن کاملCancellous Screws Are Biomechanically Superior to Cortical Screws in Metaphyseal Bone.
Cancellous screws are designed to optimize fixation in metaphyseal bone environments; however, certain clinical situations may require the substitution of cortical screws for use in cancellous bone, such as anatomic constraints, fragment size, or available instrumentation. This study compares the biomechanical properties of commercially available cortical and cancellous screw designs in a synth...
متن کاملThe biomechanics of interference screw fixation of patellar tendon anterior cruciate ligament grafts.
Twenty-seven paired human cadaveric knee specimens were used to determine the effect of surgical technique and various interference screw parameters on the pullout strength of patellar tendon femoral bone blocks. The study compared the fixation strength of endoscopically inserted and conventional "rear-entry" screws of different diameters and lengths. In all tests the most frequent mode of fail...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of orthopaedic trauma
دوره 24 6 شماره
صفحات -
تاریخ انتشار 2010