Reduction of Endotracheal Tube Connector Dead Space Improves Ventilation: A Bench Test on a Model Lung Simulating an Extremely Low Birth Weight Neonate.

نویسنده

  • Vadim A Ivanov
چکیده

BACKGROUND The reduction of instrumental dead space is a recognized approach to preventing ventilation-induced lung injury in premature infants. However, there are no published data regarding the effectiveness of instrumental dead-space reduction in endotracheal tube (ETT) connectors. We tested the impact of the Y-piece/ETT connector pairs with reduced instrumental dead space on CO2 elimination in a model of the premature neonate lung. METHODS The standard ETT connector was compared with a low-dead-space ETT connector and with a standard connector equipped with an insert. We compared the setups by measuring the CO2 elimination rate in an artificial lung ventilated via the connectors. The lung was connected to a ventilator via a standard circuit, a 2.5-mm ETT, and one of the connectors under investigation. The ventilator was run in volume-controlled continuous mandatory ventilation mode. RESULTS The low-dead-space ETT connector/Y-piece and insert-equipped standard connector/Y-piece pairs had instrumental dead space reduced by 36 and 67%, respectively. With set tidal volumes (VT) of 2.5, 5, and 10 mL, in comparison with the standard ETT connector, the low-dead-space connector reduced CO2 elimination time by 4.5% (P < .05), 4.4% (P < .01), and 7.1% (not significant), respectively. The insert-equipped standard connector reduced CO2 elimination time by 13.5, 25.1, and 16.1% (all P < .01). The low-dead-space connector increased inspiratory resistance by 17.8% (P < .01), 9.6% (P < .05), and 5.0% (not significant); the insert-equipped standard connector increased inspiratory resistance by 9.1, 8.4, and 5.9% (all not significant). The low-dead-space connector decreased expiratory resistance by 6.8% (P < .01) and 1.8% (not significant) and increased it by 1.4% (not significant); the insert-equipped standard connector decreased expiratory resistance by 1.5 and 1% and increased it by 1% (all not significant). The low-dead-space connector increased work of breathing by 4.7% (P < .01), 3.8% (P < .01), and 2.5% (not significant); the insert-equipped standard connector increased it by 0.8% (not significant), 2.5% (P < .01), and 2.8% (P < .01). CONCLUSIONS Both methods of instrumental dead-space reduction led to improvements in artificial lung ventilation. Negative effects on resistance and work of breathing appeared minimal. Further testing in vivo should be performed to confirm the lung model results and, if successful, translated into clinical practice.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of Umbilical Cord Nucleated Red Blood Cell Count in Discharged and Dead Very Low Birth Weight Infants

Background: Predictive value of the umbilical nucleated red blood cells (NRBCs) at birth for neonatal outcomes has been assessed. Hence, the present study was conducted to compare NRBC between discharged and dead neonates.Methods: This cross-sectional study was performed on preterm newborns with a birth weight of < 1,500 g admitted to the Neonatal Intensive Care Unit of Ghaem Hospital, Ma...

متن کامل

Development of pulmonary air leak in an extremely-low-birth-weight infant without mechanical ventilation: a case report

A 34-year-old woman at 25 weeks 3 days was diagnosed with preterm labor and underwent an emergency cesarean section. The neonate did not cry or show any activity. The heart rate was 80 beats/min and the oxygen saturation on pulse oximetry (SpO2) was 77%. Immediately, positive pressure ventilation was delivered by T piece resuscitator, and then anesthesiologist performed endotracheal intubation....

متن کامل

میزان کلونیزاسیون باکتریال کانکتور لوله تراشه پس از24 ، 48 و 72ساعت استفاده مکرر در اتاق عمل

Background: Colonization of bacteria on the equipments in touch with airway and respiratory mucosa may cause respiratory infections. According to available protocols, endotracheal tube connectors should be disposable, but because of financial issues many hospitals use it more than once. In the present study, the colonization rate of endotracheal tube connectors after 24, 48 and 72 hours of admi...

متن کامل

Ventilatory management in extremely low birth weight infants

The improvement in survival in premature infants associated with the evolution of mechanical ventilation has been accompanied by an increase in ventilator induced lung injury. High frequency ventilation has been shown to reduce the incidence of ventilator induced lung injury and hence chronic lung disease in the very low birth weight infant. The evolution in understanding how to best use high f...

متن کامل

Sound Analysis in an In Vitro Endotracheal Tube Model

BACKGROUND/AIMS Complete endotracheal tube obstruction is a medical emergency, and partial occlusion causes increased breathing rates and failure to wean off mechanical ventilation. Partial occlusion may be underestimated due to the lack of proper detection methods. We tested whether the sound of an endotracheal tube could be used to detect an endotracheal tube obstruction using an in vitro mod...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 61 2  شماره 

صفحات  -

تاریخ انتشار 2016