Efficacy of halogen photopolymerization units in private dental offices in Toronto.
نویسندگان
چکیده
PURPOSE Light units used for polymerization of resin composites are subject to deterioration with age, and frequent maintenance is required to maintain optimal efficacy. This study examined the efficacy of quartz-tungsten-halogen light units in private dental offices in Toronto for polymerization of resin composites. MATERIALS AND METHODS One hundred dental offices met all selection criteria and agreed to participate in the study. The light intensity was determined for a total of 214 light units. Disk-shaped specimens, 2.5 mm thick, were made from 2 resin composites (Charisma, Heraeus Kulzer; Point 4, Kerr Corp) and were subjected to photopolymerization: Charisma for 20 seconds (99 units) and Point 4 for 20 and 40 seconds (all 214 units). Knoop hardness values for the upper and lower surfaces of each specimen were determined, and relative hardness values (hardness of lower surface/hardness of upper surface x 100) were calculated. Data were analyzed using descriptive statistics, t-tests, 1-way analysis of variance, and simple and multiple linear regression (a = 0.05). RESULTS The light intensity of the individual units varied widely, from 120 to 1,000 mW/cm2. Surface hardness and relative hardness were significantly (p < 0.05) and positively associated with light intensity, and wide ranges in surface hardness and relative hardness values were observed. Mean relative hardness ranged from 34.8% to 57.7%. CONCLUSIONS Light polymerization units in private dental offices displayed a wide range in light intensity, and many had below-recommended levels. Of the resin composite specimens polymerized for 40 seconds with each of the 214 light units, only 10% reached the desired relative hardness of at least 80%. A positive linear relationship was found between light intensity and relative hardness. Increased exposure time resulted in a significant increase in relative hardness. Also, relative hardness was found to be dependent on the brand of composite material used. Dentists should regularly monitor the condition of light units and replace deteriorating parts.
منابع مشابه
آلودگی میکروبی و قارچی هوای کلینیکها و مطبهای دندانپزشکی شهر کرمان
Introduction: There is a high concern regarding air contamination in the dental clinics and offices. For the reason that the most risking factor is the presence of floating particles in the clinics containing different disease-causing microorganisms and in order to provide a better strategy to improve the condition, the present article takes into consideration the study of the air inside the me...
متن کاملبررسی کارائی دستگاههای لایتکیور در مطبهای دندانپزشکی خصوصی شهر تهران در سال 1384
Background and Aim: Light cured composites and other restorative materials are quite common in dentistry today. Successful restorations are dependent on efficiency of curing light unit, eg. the intensity of emitted light and its wave length. The aim of this study was to evaluate the efficiency units of curing, in private dental offices in Tehran. Materials and Methods: In this descriptive cross...
متن کاملQuantity and Quality Analysis and Management of Dental Solid Waste Produced in Private Office in Urmia
Objective: Dental solid wastes often contain hazardous substances. The first step of management of dental solid wastes includes detection and classification of these substances. This study aimed to qualitatively and quantitatively assess the management of solid wastes by dental offices in Urmia city in 2013. Methods: This descriptive study was conducted on all private dental offices in Urmia...
متن کاملComparative evaluation of the effect of Light Emitting Diode (LED) and Quartz Tungsten Halogen (QTH) light curing units on color stability of Filtek Z350 XT
Introduction:Discoloration of the resin-based composites is a common problem in restorative dentistry. There are many factors associated with the discoloration of dental materials in the oral environment. The purpose of this study was to evaluate the color changes in a nano-composite cured with a quartz-tungsten-halogen (QTH) and light emitting diode (LED) unit. Methods:80 disk-shaped specim...
متن کاملTime dependent accuracy of dental radiometers.
Inadequate intensity of the polymerization light source can compromise the quality and longevity of dental composite restorations. In order to maintain optimal strength of polymerization devices, regular control of polymerization units is necessary. The aim of this study was to compare the accuracy of two radiometers in the measurement of light intensity of photopolymerization devices concernin...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal
دوره 71 8 شماره
صفحات -
تاریخ انتشار 2005