Evaluation of a Plasma hCG Method for Point of Care Testing with the Aim of Shortening Test-Turnaround-Times

نویسندگان

  • Anna-Karin Wikström
  • Magnus Hagmar
  • Göran Ronquist
  • Anders Larsson
چکیده

Objective: To examine the correlation between plasma hCG results obtained with the new i-STAT® hCG point of care test with those concomitantly obtained from the central hospital laboratory utilizing the same patient samples. Methods: Prospective cross-sectional laboratory test evaluation. We compared plasma hCG results obtained with the i-STAT® hCG test (Abbott Point of Care, Princeton, NJ, USA) with Architect Ci8200 (Abbott Laboratories, Abbott Park, IL, USA). We also calculated the total coefficient of variation (CV) for the i-STAT® method. Results: The two methods showed a good linear correlation (R2 = 0.994; slope 1.03) and CV for the i-STAT® method was 2.1% 5.2%. Conclusion: We suggest that the i-STAT® hCG blood assay could be used as a complement to urine hCG assays in clinical situations when rapid test results are needed and urine is not available.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Lean six sigma process improvement in specimen receiving to improve stat chemistry turnaround times

  Objective: As a consequence of stat turnaround times (TATs) chronically exceeding 60 minutes, our laboratory was facing pressure to divert limited resources toward the implementation of an emergency department satellite laboratory. Peer-reviewed literature in clinical laboratory quality assurance and improvement indicates that between 60-70% of errors occur at the pre-analytical level.  Thus...

متن کامل

Comparison of Result Times Between Urine and Whole Blood Point-of-care Pregnancy Testing

INTRODUCTION Point-of-care (POC) pregnancy testing is commonly performed in the emergency department (ED). One prior study demonstrated equivalent accuracy between urine and whole blood for one common brand of POC pregnancy testing. Our study sought to determine the difference in result times when comparing whole blood versus urine for the same brand of POC pregnancy testing. METHODS We condu...

متن کامل

Evaluation of the Effect of Acupressure, LV3 Point on Maternal Physiological Parameters and Non-Stress Testing of the Fetus

Introduction: The mother's vital signs change in the face of stressful phenomena Acupressure improves vital function by secreting enkephalins. The aim of this study was to determine the effect of acupressure, LV3 point on maternal physiological parameters and non-stress testing of the fetus. Methods: A clinical trial study was performed on 84 primiparous mothers. The study tool was a questionn...

متن کامل

Exploration of the Customized Fixtures for the Evaluation of Three-point Bending Strength of Dental Resin Composites

Introduction: This study aimed to devise customized fixtures for the evaluation of three-point bending strength (TPBS) of resin-based dental composites (RBCs). Materials and Methods: A cube-shaped jig made out of wood with dimensions of 105×105×101 mm was prepared in this study. A 20-mm-diameter hole was made in the center of the wooden jig. In addition, a stai...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2015