Evaluation of an Electrochemical Point‐of‐Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle
نویسندگان
چکیده
BACKGROUND The Precision Xtra(®) meter is a promising low cost electrochemical point-of-care unit for measuring blood glucose concentration ([gluc]) in cattle blood. The meter uses an algorithm that assumes the intra-erythrocyte [gluc] equals the plasma [gluc] on a molal basis, and that the hematocrit is similar in humans and cattle. OBJECTIVES The primary objective was to determine the accuracy of the meter for measuring plasma [gluc] in dairy cattle. Secondary objectives were to characterize the influence of hematocrit and sample temperature on the measured value for [gluc]. ANIMALS A total of 106 periparturient Holstein-Friesian cattle. METHODS Blood and plasma samples (1,109) were obtained and Deming regression and Bland-Altman plots were used to determine the accuracy of the meter against the reference method (plasma hexokinase assay). Multivariable regression and linear regression were used to determine the effect of hematocrit and sample temperature on the plasma [gluc] measured by the meter. RESULTS Intra-erythrocyte [gluc] was 18% of plasma [gluc] on a molar basis. Sample temperature had a significant linear effect on plasma [gluc] as measured by the meter for 3/5 plasma samples when measured [gluc] > 160 mg/dL. CONCLUSIONS AND CLINICAL IMPORTANCE The meter utilizes an algorithm that is optimized for human blood and is inaccurate when applied to bovine blood. Until a cattle-specific algorithm is developed, we recommend using plasma as the analyte instead of blood and calculating plasma [gluc] using the equation: [gluc] = 0.66 × [gluc]p-meter + 15, where [gluc]p-meter is the value reported by the meter. If blood is measured, then we recommend using the equation: [gluc] = 0.90 × [gluc]b-meter + 15.
منابع مشابه
Evaluation of a point-of-care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows.
OBJECTIVES To evaluate and validate a hand-held electrochemical meter (Precision Xtra®) as a screening test for subclinical ketosis and hypoglycaemia in lactating dairy cattle. DESIGN Method comparison study using a convenience sample. PROCEDURE Blood samples were collected into plain tubes from the coccygeal vessels of 181 Holstein cows at 2-4 weeks of lactation during summer in Iran. Bloo...
متن کاملP-64: The Relationship between Polymorphism in Gene of Insulin-Like Growth Factor-I and The Serum Periparturient Concentration in Holstein Dairy Cows
Background: One of the most important metabolic factors affecting the reproductive activity is insulin-like growth factor-I (IGFI) concentration changes after calving. Recently, Maj et al. (2008) discovered a significant association between the IGF-I genotypes based on the 5'-untranslatedregion (5'-UTR) of IGF-I gene and the IGF-I blood level. The objective of this study is to investi...
متن کاملAssociation of Kappa-Casein Gene Polymorphism with some Biochemical Blood Indicators in Guilan Native Cattle of Iran (Bos indicus)
Evaluation of the blood and milk biochemical parameters always is considered as effective key factor for animal health and dairy products. In order to assess the association of kappa-casein (K-CN) genetic variants with blood biochemical indicators, blood samples were randomly and individually collected from 126 Iranian Guilan native cattle. Blood plasma was used to measure blood glucose, urea, ...
متن کاملArteriographic Evaluation of Laminitis Digits in the Hind Limbs of Dairy Cattle
Objective: Laminitis is one of the main causes of lameness in dairy cattle. In this situation the corium blood circulation is disrupted and the production of healthy horny hoof wall is reduced. The purpose of this study was to evaluate the arteriographic pattern of the digital arterial branches in the laminitic digits and to compare them with the normal digits. Design: Original study. Animals...
متن کاملAnalytical Performance and Quality Control of a Glucose Monitor System
Background and Objective: The reliability and validity of monitors for self-monitoring of blood glucose are debated. We evaluated the analytical performance of Accu-check Active (Boehringer Mannheim, Roche) which is one of the most commonly used monitors in Iran. Material and Methods: We compared the monitor readings with the reference values by percentage of values within...
متن کامل