Solid water phantom heat conduction: Heating and cooling rates

نویسندگان

  • Martin J. Butson
  • Tsang Cheung
  • Peter K. N. Yu
چکیده

Solid water is often the phantom material of choice for dosimetry procedures in radiotherapy high-energy X-ray and electron beam radiation calibration and quality assurance. This note investigates variation in heat conduction that can occur for a common commercially available solid water stack phantom when a temperature differential occurs between the phantom and ambient temperature. These variations in temperature can then affect radiation measurements and thus the accuracy of radiation dosimetry. In this manuscript, we aim to investigate the variations in temperature which can occur in radiation measurement incorporated (RMI) solid water phantoms, their thermal properties and the effects on radiation dosimetry which can occur because of temperature differentials. Results have shown that the rate of temperature change at a phantom center is a complex function but appears relatively proportional to the surface area of the phantom in normal clinical usage. It is also dependent on the thermal conductivity of any material in contact with the phantom; and the nature of the phantom construction, i.e., the number and thickness of slices within the phantom. A thermal time constant of approximately 20 min was measured for a 2-cm solid water phantom slice when located on a steel workbench in comparison to 60 min when located on a wooden workbench (linac couch insert). It is found that for larger solid water stack phantoms, a transient (within 1 degrees C) thermal equilibrium exists at the center for up to 2 h, before the temperature begins to change. This is assumed to be due to the insulating properties of multiple slices within the stack, whereby very small air spaces are introduced inhibiting the heat conduction through the phantom material. It is therefore recommended that the solid water/phantom material is kept within the treatment room for closest thermal accuracy conditions or at least placed within the room approximately 10 h before dosimetry measurements. If these options are not available, a standard linear interpolation method for calculation of temperature should be used to minimize uncertainty of temperature measurements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Entropy Generation Analysis of Transient Heat Conduction in a Solid Slab with Fixed Temperature Boundary Conditions

Analysis of entropy generation rate for transient heat conduction, taking place in homogeneous solid slab with fixed temperature boundary conditions, is presented. The initial condition is defined as a constant temperature. The exact solution is solved and used to analyze the local and total entropy generation rates. Two cases, heating and cooling processes, are considered in this study. It is ...

متن کامل

Modeling and Analysis of a Solid Oxide Fuel Cell Based Trigeneration System with an Oxygenated Fuel by Using an Exergoeconomic Methodology for Power, Heating and Cooling Production

In the present study, thermo-economic analysis of a combined solid oxide fuel cell (SOFC) with a gas turbine, a generator-absorber heat exchanger (GAX) and heating process heat exchanger for heating, cooling and power production as a tri-generation system is conducted. Also, an external steam reformer is applied to convert methanol as oxygenated fuel to hydrogen for the electrochemical process ...

متن کامل

CFD Modelling of an Underground Water Tank Heat Storage System

In this work, CFD is used to simulate the flow in an existing thermal labyrinth located at the basement of a hospital building. This labyrinth is an underground channel equipped with water tanks to increase the thermal mass and reduce the heating and cooling demand of the building. Thermal energy storage in the water tanks is a complex heat transfer phenomena involving turbulent forced convecti...

متن کامل

What Spectral Methods Can Tell Us about Heat Transfer Effects in Water Calorimeters

In water calorimetry, numerous ancillary effects – such as radiation-induced chemical reactions, scattering and excess heat from nonwater materials, and dose nonuniformities within the phantom – complicate the determination of absorbed dose from measurements of radiation-induced heating. Corrections for thermal transport due to excess heat and dose nonuniformities can be difficult to assess bec...

متن کامل

Heat transfer in a microvascular network: the effect of heart rate on heating and cooling in reptiles (Pogona barbata and Varanus varius).

Thermally-induced changes in heart rate and blood flow in reptiles are believed to be of selective advantage by allowing animal to exert some control over rates of heating and cooling. This notion has become one of the principal paradigms in reptilian thermal physiology. However, the functional significance of changes in heart rate is unclear, because the effect of heart rate and blood flow on ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 33  شماره 

صفحات  -

تاریخ انتشار 2008